DO Tip (7544)

Instruction Manual

Thank you for purchasing the HORIBA DO electrode. Read
this manual carefully before using the electrode.

*

"DO" is an abbreviation of Dissolved Oxygen.

Conformable standards

This equipment conforms to the following standards:

Authorised representative in EU and UK

HORIBA Europe GmbH
Hans-Mess-Str.6, D-61440 Oberursel, Germany

HORIBA UK Limited
Kyoto Close, Moulton Park, Northampton, NN3 6FL, UK

Cautions during handling

An extremely thin membrane is stretched across the DO
tip and DO responsive part. Do not allow the membrane
to be in contact with any hard surface and do not exert
undue force on the diaphragm.

When attaching the DO tip to the electrode body, check
to make sure that O-ring is attached. Attach the DO tip
securely to the electrode body.

DO tips that are no longer usable should be disposed of
as industrial waste.

Packaged contents

Specifications

RoHS: EN 

IEC 

63000

9. Monitoring and control instruments including 
industrial monitoring and control instruments

RoHS: BS 

EN 

IEC 

63000

9. Monitoring and control instruments including 
industrial monitoring and control instruments

WARNING

Chemical solution

Highly concentrated potassium hydroxide (KOH) is used in the 
internal solution of the electrode.
If the internal solution in the electrode comes in contact with your 
hands or skin, wash immediately with water. If the internal solution 
comes in contact with your eyes, flush immediately with large 
amounts of water and seek treatment by a physician. Follow the 
instructions on the associated Safety Data Sheet (SDS).

Name

Q'ty

DO tip (7544)

1 pc

Instruction manual

1 copy

Packing box

1 pc

Model

7544

Measurement principles

Membrane galvanic cell

Measurement range

0 mg/L to 20.00 mg/L

Usable temp. range

0°C to 50°C

Materials

PPS

*

, Sn, Ag

Ext. diam. of wetted part

15.6 mm

*

: Wetted material

Storage

The electrode should remain to be connected to the
meter during storage.

To store the electrode without the meter, follow the
below procedure.

1. Remove the electrode form the connector on the meter.
2. Clean the electrode well with tap water, then dry it with a

soft cloth.

3. Remove the DO tip from the holder.
4. Put the short socket on the DO tip.
5. Store the DO tip in a closed container in a cool, dark

location with an oxygen scavenger.

Note

Avoid storing the electrode in hot place or places with high
humidity. Sore the electrode indoors and out of direct
sunlight.

The color of the internal solution may change depending on
storage conditions but does not affect performances of the
product.

Installation procedure

1. Remove the DO tip (7544) form the package.
2. Remove the short socket from the DO tip.

3. Attach the sensor adapter to the holder when measuring

in BOD bottle.

4. Attached the DO tip to the holder.

5. Put the stirrer into the stirrer holder.

6. Attach the stirrer holder to DO tip.

Note

Do not dispose of the short socket that was removed. It is
necessary when storing the electrode.

When attaching the DO tip, make sure that the O-ring is
attached to the DO tip, and screw the DO tip on firmly, until
it stops.

Disposal method

When disposing of the electrode, follow the related laws
and/or regulations of your country/region.

Short socket

DO tip

O-ring

Holder

Sensor adapter

Holder

DO tip

Stirrer holder

Stirrer

Stirrer holder

DO tip