Optical DO Sensor

300-D-2 / 300-D-5

*”DO” is an abbreviation of Dissolved Oxygen.

Warranty and Responsibility

HORIBA Advanced Techno Co., Ltd. warrants DO sensor shall be

free from defects in material and workmanship and agrees to repair

or replace free of charge, at option of HORIBA Advanced Techno

Co., Ltd., any malfunctioned or damaged product attributable to

responsibility of HORIBA Advanced Techno Co., Ltd. for a period of

Three (3) years from the delivery unless otherwise agreed in a

written statement. Sensor cap is eligible for a replacement against

incipient defects within Two (2) years from the delivery. In any one of

the following cases, none of the warranties set forth herein shall be

extended:

Any malfunction or damage attributable to improper operation

Any malfunction attributable to repair or modification by any

person not authorized by HORIBA Advanced Techno Co., Ltd.

Any malfunction or damage attributable to the use in an

environment not specified in this manual

Any malfunction or damage attributable to violation of the

instructions in this manual or operations in the manner not

specified in this manual

Any malfunction or damage attributable to any cause or causes

beyond the reasonable control of HORIBA Advanced Techno

Co., Ltd. such as natural disasters

Any deterioration in appearance attributable to corrosion, rust,

and so on

HORIBA Advanced Techno Co., Ltd. SHALL NOT BE LIABLE FOR

ANY DAMAGES RESULTING FROM ANY MALFUNCTIONS OF

THE PRODUCT, ANY ERASURE OF DATA, OR ANY OTHER USES

OF THE PRODUCT. 

Conformable standards

This equipment conforms to the following standards:

Authorised representative in EU

HORIBA Europe GmbH

Hans-Mess-Str.6, D-61440 Oberursel, Germany

Cautions during handling

Do not allow the sensor to be in contact with any hard surface.

The surfaces of the sensor cap is covered with

fluorescent materials. Do not allow the sensor cap to come in

contact with dirt and operate carefully when immersing into a

solution.

Avoid continuous measurements in solutions with strong acids,

solutions with strong alkalis and solutions with organic solvents.

Electromagnetic interferences may cause malfunctions in the

DO sensor.

Packaged contents

Specifications and names of parts

Specifications

RoHS:

EN50581
9. Monitoring and control instruments

DO sensor (with sensor cap)

1 pc.

Calibration bottle

1 pc.

Protective guard

1 pc.

Instruction manual

1 copy

Model

300-D-2/300-D-5

Measurement method

Fluorescence quenching

Measurement range

0 to 20 mg/L

Accuracy  (DO)

±2 % *

Temperature range

0 to 50 °C

Accuracy (Temperature)

±0.3 °C 

Response time

30 s (t :90)

Watted materials

DO sensor: PC/ABS

Sensor cap: PMMA

Protective guard: SUS316L

DO sensor size

16×200

Lead cable length

2 m (300-D-2)
5 m (300-D-5)

Waterproof

IP67

*For a calibration perfomed in saturated water

Names of parts

Note

The sensor cap is delivered equiped with the DO sensor.

Perform a calibration before any measurement.
The calibration method is indicated in the instruction
manual of the meter. 

It is recommended to make a calibration in liquids for a
better accuracy during measurements.

Make sure there are no air bubbles visible on the surface of
the cap during a measurement. 

After finishing a measurement, clean the sensor by using a
soft cloth. Take care not to scratch the sensor cap. After
cleaning, rinse the sensor cap with distilled water.

Store the DO sensor with the sensor cap and the calibration
bottle in a cool, dark location. The sponge inside the
calibration bottle should be kept wetted.

Sensor cap replacement instruction procedure

1. Remove the old sensor cap from the DO sensor.

2. Tighten the new sensor cap on the DO sensor.

3. Input the membrane parameter indicated on the sensor cap in

the meter.

4. Perform calibration.

The instruction procedure has been completed.

Note

The membrane parameter of the sensor cap delivered is set
in the meter.

The setting method of the membrane parameter is indicated
in the instruction manual of the meter.

DO sensor

Calibration bottle

Sensor cap

Protective cap

(300-D-M)

DO sensor

Sensor cap

xxxx

Sensor cap

DO sensor

Membrane parameter

Enlarged