Calibrating temperature sensor
40
HORIBA
●
Calibrating temperature sensor
The temperature sensor or temperature compensation electrode in the combination
electrode has ±1°C accuracy without calibration. You can use a known temperature
solution to calibrate the temperature sensor.
1. Immerse the electrode into the solution until the temperature sensor is
immersed.
2. Press the
key to enter the setting mode.
3. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
4. Press the
keys to select the "TEMP" (temperature calibration setting) and
then press the
key.
5. Enter the set temperature and press the
key.
The temperature sensor is calibrated.
To return to the setting mode, press the
key.
Note
When initializing temperature calibration data, all settings need to be initialized. Perform
initialization by referring to “ Resetting to factory default settings ” (page 43). When
initialization is performed, all saved data is deleted. Copy or transfer necessary data to a
PC for storage.