When an error message appears
44
●
ERR No.0008
Calibration interval error
・
More than the set number of days has elapsed since calibration was last performed.
・
The calibration interval alarm is "ON" and calibration has not been performed.
●
ERR No.0010
Memory full
Attempted to save more than 1000 items of data.
Cause
How to solve problem
Calibration has not been
performed for the set number of
days of the calibration interval or
longer.
Perform calibration.
The calibration interval alarm is
"ON" and calibration has not
been performed.
Perform calibration.
Cause
How to solve problem
Saving more than 1000 items of
data is attempted.
The maximum number of savable items of data is 1000.
Copy necessary data and delete the data from the
memory (“ Deleting all saved data ” (page 30)).