pH calibration
11
■
Calibration
This section describes the basic calibration method using LAQUA PD200 handheld meter, pH
electrode and DO electrode.
●
pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration, follow the
procedure detailed below.
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the PD meter and plug in the pH electrode.
• Prepare buffer solution required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the buffer solution.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH7 first.
• Default buffer setup is
BUFF USA
. If you want to change to
BUFF NIST
or
BUFF DIN
,
refer to “P 1.1 Buffer selection” on page 26.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to “P 1.3 Erase calibration data” on page 28.
Note
Tip
Summary of Contents for LAQUA-PD210
Page 1: ......
Page 2: ......
Page 49: ...Data set up 35 DO mode...
Page 51: ...Data set up 37 DO mode...
Page 53: ...Data set up 39 DO mode...
Page 56: ...General set up 42 DO mode...
Page 58: ...General set up 44 DO mode...
Page 60: ...General set up 46 DO mode...
Page 62: ...General set up 48 DO mode...
Page 65: ...Clock set up 51 DO mode...
Page 67: ...Clock set up 53 DO mode...
Page 86: ...Specifications 67...
Page 88: ......