Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
If the ion selective membrane is rubbed severely to remove
the moisture, the ion selective membrane may tear or static
electricity may be generated, resulting in a measurement
errors. Press the cloth against the ion selective membrane
lightly so as to soak up the water.
When disposing of the standard solution, and paper or cloth
soaked in the standard solution, follow the local rules and
regulations.