U408_flowthermNT_B_e_160420
31/35
www.hoentzsch.com
When using rechargeable batteries, "rechargeable" must be selected in the settings (5.3.22) other-
wise the battery capacity display will be incorrect!
Device profiles:
5.3.26
Device profiles
Menu -> Profiles
The entire parameter inputs can be saved in the profiles under a freely definable name with up to 8 char-
acters and can subsequently be reloaded.
For example, all parameter inputs for a specific sensor can be saved to a profile or also to an appointed
measuring point.
Up to 100 different profiles can be saved. Available profile storage locations are marked with an
*
after
the profile name. Profiles cannot be deleted but may be overwritten.
Returning to the factory settings (5.3.23) has no impact on the saved profiles.
Selection:
Load:
=
loading a saved profile
by selecting from the list
Save:
=
saving the active parameter settings
to a profile by selecting from the
list in an available profile storage field and input of a new name or in
an already occupied profile storage location by overwriting and changing
or retaining the name
Keys:
5.3.27
Key F3 - LM-Start
Funktion key for operating long-term measurement (see 5.3.21)
5.3.28
Key 4 Switch off / OFF (reset)
Funktion key for switching off the device; active in all menus except during long-term measurement.
If the device no longer reacts to keypad entry, hold this key down for longer than 10 seconds to switch off
and reset.