______________________________________________________________________________________
U447_ExactSonicP_B_e_220628
17/18
5
Troubleshooting
Error
Reason
Troubleshooting
Device shows no
function
Power supply insufficient
Check power supply and connector
Elektronics defective
Return to the factory
No measured value
Sensor dirty
Check and clean the sensor
Incorrect measurement
settings
Comparison and correction of the settings ac-
cording to the specifications in the Technical
Data Sheet
No display of the
absolute pressure
No connection
Check cable and connector on pressure sensor
Elektronics or display
defective
Return to the factory
No indication in the
display
Power supply insufficient
Check power supply and connector, check sta-
tus LED (see chapter 3.5)
Elektronics or display
defective
Return to the factory
Display cannot be
operated
User interface locked
Unlock the user interface according to chapter
3.1
Incorrect value at
analog output
Incorrect settings for analog
output
Check the settings according to chapter 3.3 or
chapter 4.4.5
Measured value too
low
Sensor dirty
Check and clean the sensor
Incorrect measurement
settings
Comparison and correction of the settings ac-
cording to the specifications in the Technical
Data Sheet
Measured value too
high
Incorrect measurement
settings
Comparison and correction of the settings ac-
cording to the specifications in the Technical
Data Sheet
No TCP communica-
tion
No connection
Check cables and connectors, check accessibil-
ity via ping test
IP address set incorrectly
Check the settings according to chapter 3.3 or
chapter 4.4.5
Port set incorrectly
Check the settings according to chapter 3.3 or
chapter 4.4.5