______________________________________________________________________________________
U447_ExactSonicP_B_e_220628
11/18
3.4
Tab Info
Task: Dispal of device information
The tab Info contains the device name, the version number, the serial number, the device time and the
contact details of Höntzsch GmbH & Co KG.
The device's time is set automatically when there is an internet connection. If the network in which the
ExactSonic P is integrated does not have access to the Internet or the device has no authorization, the
date and time can be set using the AK protocol (see AK command „ESYT“ in chapter 4.4.5).
3.5
Status LED
Task: Display of various statuses independent of the display
Colour/state
Description
Off
Permanent: Power supply interrupted
Red
1 x short flash, 2 seconds off: error in evaluation of ultrasonic data
2 x short flash, 2 seconds off: error in user-interface
3 x short flash, 2 seconds off: error in analog output
4 x short flash, 2 seconds off: error in temperature-, pressure- or humidity sensor
5 x short flash, 2 seconds off: calculation error
Green
Permanent: user interface started, device in operation
Blue
Permanent: boot process
Yellow
Permanent: device is busy