Introduction 

The GasAlertMicro 5/PID/IR Battery Charger (“the charger”) charges the 
GasAlertMicro 5, GasAlertMicro 5 PID, and the GasAlertMicro 5 IR 
rechargeable battery packs. 
 

Standard GasAlertMicro 5/PID/IR Battery Charger kit includes: 

 

GasAlertMicro 5/PID/IR Battery Charger unit 

 

Universal power adapter 

 Power 

cord 

 

Wall mounting bracket 

 Instruction 

sheet 

Table 1. Order Number 

Model No. 

Description 

Qty 

M5-C01

*

 

 

 

Cradle 

charger

1

*

Add suffix (-UK) for United Kingdom mains plug, (-EU) for European mains 

plug, (-AU) for Australian mains plug. 

 

 

 

 

 

 

D5688/2 
iERP: 124011 
© BW Technologies 2006, all rights reserved. Printed in Canada. 
All product names are trademarks of their respective companies.

a

 

Safety Information - Read First 

 

For indoor use only. 

 

Warning

: Substitution of components may impair Intrinsic Safety of the 

detector being charged. 

 

Do not charge the battery pack with any other unapproved charger. Do not 
use the charger, change the batteries, or charge the batteries in a 
hazardous location. Failure to observe these precautions will impair the 
Intrinsic Safety of the detector and can lead to fire and/or explosion. 

 

Do not expose the charger to electrical

 

shock or severe continuous 

mechanical shock.

 

 

Ensure the battery pack is locked in place prior to operating the detector. 

 

If the charger is damaged or parts are missing, contact  

BW Technologies by Honeywell

 immediately. 

 

To eliminate the risk of electrical shock, unplug the charger whenever 
cleaning or other maintenance is required. 

 

Avoid touching the charging cradle contact pins. 

 

Use only BW approved batteries. Do not use alkaline or other non-
rechargeable batteries with this charger. 

 

The detector must be deactivated before charging. 

 

Do not calibrate the detector during or immediately after charging is 
complete. 

Table 2. International Symbols 

Symbol 

 

Meaning

n

Approved to both U.S. and Canadian Standards by the CSA 
International 

a

Caution – risk of danger 

dc current 

Parts of the GasAlertMicro 5/PID/IR Battery Charger 

 

 

Retaining hooks 

Charging cradle 

Battery connectors (contact pins) 

Charger status LED 

Mounting access hole 

Retaining base 

 

Summary of Contents for M5-C01 Series

Page 1: ...ttery pack with any other unapproved charger Do not use the charger change the batteries or charge the batteries in a hazardous location Failure to observe these precautions will impair the Intrinsic Safety of the detector and can lead to fire and or explosion Do not expose the charger to electrical shock or severe continuous mechanical shock Ensure the battery pack is locked in place prior to ope...

Page 2: ...r emits a low battery alarm charge the battery immediately When charging a new battery for the first time fully charge the battery pack before use approximately 3 5 hours Note If a fully charged battery is inserted into the charger the charger status LED lights red for 6 to 10 minutes and then changes to green does not damage the battery The detector can remain in the cradle after charging is comp...

Page 3: ...the battery pack refer to Figure 5 and complete the following 1 Deactivate the detector 2 Open the latch on the bottom of the detector 3 Remove the battery pack by lifting the bottom of the pack away from the detector 4 Insert the new battery pack Close and press the latch until the release tabs click Figure 5 Removing the Battery Pack Charging the Battery Pack without the Detector To charge the b...

Page 4: ...oneywell The charge LED does not activate when the detector is inserted Damaged or defective battery pack Contact BW Technologies by Honeywell There is no audible beep Internal damage to battery or charger Contact BW Technologies by Honeywell The battery is above or below the operating temperature Leave the battery in the charger When the battery returns to room temperature charging will begin The...

Page 5: ...cts to the original buyer BW s warranty obligation is limited at BW s option to refund of the purchase price repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period In no event shall BW s liability hereunder exceed the purchase price actually paid by the buyer for the Product This warranty does not include a fuses disposable batter...

Page 6: ...ing the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 1245 600560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...