Honeywell HP970A Service Data Download Page 3

 5 AND HP9725 PNEUMATIC 

 LOCK ADJUSTMENT

Set the minimum and maximum 

 as follows:

 Loosen the 

 lock screw (Fig. 4).

 Move the 

 locking stops to the desired position.

 Tighten the 

 lock screw.

 Check the stop positioning by turning the 

adjustment.

CALIBRATION CHECK

Measure the rh with an accurate rh measuring device.

Turn the 

 adjustment until the 

 indicator

reads 

 percent below the indicated rh and allow the

Humidistat to build up BLP.

Turn the 

 indicator up slowly using the 

adjustment.

if the Humidistat bleeds off at 13 percent rh of the

indicated humidity, no further calibration is necessary. If

the Humidistat does not bleed off properly, proceed with

RECALIBRATION.

 and 

Measure the rh with an accurate rh measuring device.

Turn the 

 adjustment until the 

 indicator

reads 10 percent above the indicated rh and allow the
Humidistat to build up BLP.

Turn the 

 indicator down slowly using the

 adjustment.

If the Humidistat bleeds off at 

 percent rh of the

indicated humidity, no further calibration is necessary. If

the Humidistat does not bleed off properly, proceed with

RECALIBRATION.

RECALIBRATION

CAUTION

These Humidistats are sensitive devices and should
be handled with extreme care during calibration. Do
not breathe directly on the element when the cover

is off.

Ensure the system is stabilized and the main air is
maintained at normal pressure between 13 and 21
psi (90 and 145 

 Factory calibration is 18 psi

(125 
Remove the Humidistat cover.
Install the Pressure Gage into the gage tap (Fig. 4)
using the Gage Adapter.
Measure the rh with an accurate rh measuring device.
Use the 

 adjustment to set the 

indicator at the actual relative humidity.

Adjust the calibration screw (Fig. 4) until the Pressure
Gage reads 0 psi (0 

Readjust the calibration screw slowly until the
Pressure Gage reads 8   psi (55 

 

Allow sufficient time for a complete response and
stabilization, then recheck pressure. The Humidistat
is now calibrated.

Remove the Pressure Gage and Gage Adapter.
NOTE: If the gage tap fails to seal itself properly

after the Pressure Gage is removed, bleed
off a small amount of air and insert Gage

Tap Plug 14002172-001 into the gage tap.

Replace the Humidistat cover.
Set the Humidistat at the desired setting.

TROUBLESHOOTING

Table 1 lists common Humidistat problems, their possible
causes, and solutions. For any problems not covered in Table
2, Troubleshooting Flowchart (Fig. 5).

Summary of Contents for HP970A

Page 1: ...tings to convert older models or competitors humidistats Figure 1 shows a typical HP970A 13 or HP972B two pipe hookup Figure 2 shows a typical HP972B one pipe hookup HP970A B HP972B MB VALVE SUPPLY N C C322 5 Fig 1 Typical HP970A 6 or HP972B Humidistat Two Pipe Hookup HP972B RESTRICTION BLUE HUMIDIFIER VALVE N C SUPPLY C3226 Fig 2 Typical HP972B Humidistat One Pipe Hookup SPECIFICATIONS Action HP9...

Page 2: ...outdoor temperature preventing condensation during extreme winter conditions The tools shown in Figure 3 and a sling psychrometer or Digital Relative Humidity Pen CCT915 are necessary to calibrate the Humidistats PEN CCT915 0 if the nylon sensing element becomes contaminated with oil grease or dust remove and clean it by meetly agitating it in an acetone solution Dry the element at room temperatur...

Page 3: ...oes not bleed off properly proceed with RECALIBRATION RECALIBRATION CAUTION These Humidistats are sensitive devices and should be handled with extreme care during calibration Do not breathe directly on the element when the cover is off Ensure the system is stabilized and the main air is maintained at normal pressure between 13 and 21 psi 90 and 145 kPa Factory calibration is 18 psi 125 kPa Remove ...

Page 4: ...HP97QA i AND HP9725 PNEUMATIC HUMIDISTATS Table 1 Common Problems Causes and Solutions 75 7195 ...

Page 5: ...HP970A BAND HP972B PNEUMATIC HUMIDISTATS RESSURE IS REDUCE 1 N O Z Z L E SEMBLY f CHECKTHE MAIN AIR PRESSURE AT THE PRV PRESSURE DROP IN THE LONG AIR Fig 5 Humidistat Troubleshooting Flowchart 75 7195 ...

Page 6: ...e flat on the hub of the setpoint cam and wheel assembly Check calibration of Humidistat and recalibrate if necessary Remove the Humidistat cover Remove the old Humidistat leaving the airhead connected to the mounting plate Discard the new mounting plate Align the port holes Fig 6 on the back of the replacement Humidistat with the air connections on the existing airhead Snap the new Humidistat ont...

Page 7: ...acement unit in reverse order Make certain that the two levers are free to move Recalibrate the Humidistat Reinstall the Humidistat cover CALIBRATION LEVER ARM7 RELIEF SPRING SPRING POST Fig 8 Back of the Humidistat Showing Relief Spring Connections SCREW __ CALIBRATION CALIBRATION LEVER HOOK r MOUNTING SCREW NO 4 x 3 16 PN HD NOZZLE FLAPPER ELEMENT CALIBRATIO ELEMENT L E V E L HOOK ASSEMBLY YOKE ...

Page 8: ...ng these parts use extreme care to prevent dirt dust or debris from entering various openings of the Humidistat base or the restrictor block Remove the humidistat cover Remove the Humidistat 3 6333 2 Fig Il Restrictor Block Assembly and Filter Cartidge Replacement Access to the restrictor block assembly and filter cartridge is from the rear of the base assembly Remove the four mounting screws from...

Page 9: ...t Covers Part Number Suffix Window insert Setpoint 20 80 RH o Window Insert 03040506070 Vhen ordermg use complete part number rncludrng three digrt suttrx 14004407 XXX Covers are paintable See HP970 72 and TP970 74 Pneumatic Sensors Humidistats and Thermostats Installation instructions 955597 for painting information ACCESSORIES Table 4 HP970A B and HP9728 Pneumatic Humidistat Accessories Parts Li...

Page 10: ...lping You Control Your World 75 7195 8 89 By using this Honeywell literature you agree that Honeywell will have no liability for any damages arising out of your use or modification to the literature You will defend and indemnify Honeywell its affiliates and subsidiaries from and against any liability cost or damages including attorneys fees arising out of or resulting from any modification to the ...

Reviews: