GENESIS DM CLINICAL GUIDE
Patient Training
P.4510EN.01
12/5/2007
52
Health Frontier ecg@home
This section explains how to program the Genesis DM monitor to prompt for an ECG recording, setup
the ecg@home device, and instruct the patient in obtaining an ECG recording. For information about
viewing ECG recordings in LifeStream, refer to your
LifeStream Clinician Guide
.
If you have any questions or concerns, call Customer Service. For repair or shipping information, refer
to the
Getting Started
section of this guide.
Assigning a COM port
To enable the monitor to prompt for and collect ECG data, you must assign a COM port. If you do not
assign a COM port, the monitor
will not prompt
for the device and it
will not collect data
from
that device even if the device is connected to the COM port.
It is also important to use an
isolated
port. Isolated ports protect patients from any potential of shock
when they use a device that is connected to the monitor.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to
Section 3 Programming the Genesis DM Monitor
for complete
programming instructions.
1.
Enter Setup Mode, scroll to the Option Menu and press
RETEST
to enter the menu.
2.
In the ECG Monitor menu, assign either COM1 or COM 2.
3.
In the ECG Time menu, set the collection frequency.
4.
Exit Setup Mode.