45
The last recorded measurement of gas temperature and is scaled to the selected
temperature units at 551.
Default: -40.00
503
Case Temperature
The last recorded measurement of the air temperature inside the recorder, considered an
ambient temperature measurement.
Default: -40.00
504
Case Maximum Temperature
The highest recorded case temperature measurement since the recorder was last reset.
Default: -40.00
505
Case Minimum Temperature
The lowest recorded case temperature measurement since the recorder was last reset.
Default: 165.00
506
P1 Pressure Calibration Zero Offset
The most recent span value for P1 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 1.0000 and recalculated on each
Pressure Span Calibration
.
Default = 0.000
507
P1 Previous Calibration Zero Offset
The previous P1 span factor recorded in item 506 and is updated on each calibration.
Default: 0.000
508
P1 Pressure Calibration Span Factor
The most recent span value factor for P1 pressure transducer and obtained during a
pressure span calibration. The value is normally very close to 1.0000 and recalculated on
each pressure span calibration.
Default = 1.0000
509
P1 Previous Calibration Span Factor
The previous P1 span factor recorded in item 508 and is updated on each calibration.
Default: 1.0000
510
P2 Calibration Zero Offset
The most recent span value for P2 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 0.0000 and recalculated on each
Pressure Span Calibration
.
Default: 0.0000
511
P2 Previous Calibration Zero Offset
The previous P2 zero offset value recorded in item 510 and is updated on each
calibration.
Default: 0.0000
Item Item Name & Description
512
P2 Pressure Calibration Span Factor
The most recent span value factor for P2 pressure transducer and obtained during a
pressure span calibration. The value is normally very close to 1.0000 and recalculated on
each pressure span calibration.
Summary of Contents for ERX
Page 68: ...68 www honeywell com ...
Page 69: ...69 ...
Page 70: ...70 www honeywell com ...
Page 71: ...71 ...
Page 72: ...72 www honeywell com ...
Page 73: ...73 ...
Page 74: ...74 www honeywell com ...
Page 75: ...75 ...
Page 76: ...76 www honeywell com ...
Page 77: ...77 ...
Page 78: ...78 www honeywell com ...
Page 79: ...79 ...
Page 80: ...80 www honeywell com ...
Page 81: ...81 ...
Page 82: ...82 www honeywell com ...
Page 83: ...83 ...