Note: BW Technologies by Honeywell recommends the sensor(s) be bump tested
immediately.
The LCD next displays the number of days remaining until the next bump test is due.
Force Bump Enabled (optional)
If the Force Bump is enabled in Safety Suite Device Configurator (SSDC) and a
sensor is overdue for a bump test, the sensor(s) must be tested to continue and
enter normal operation. The following screen displays.
Apply gas while BUMP DUE NOW displays. Verify the visual, audible, and
vibrator alarms activate. The detector must enter alarm to pass the bump test.
For complete instructions, refer to See "Bump Test" on page
If Force Bump is disabled, press
to acknowledge the warning and continue with the
startup tests.
Note: If the bump test interval field is set to 0 in Safety Suite Device Configurator (SSDC),
the startup bump test is bypassed. Refer to Bump Interval in See "Sensor Configuration" on
BW MaxXT II
24
User Manual