Maximum Operating Life

24-month Detector

2 years after activation, assuming 2 minutes 

of alarm time per day.

Note:

 

24-month H

2

S and CO detectors that 

use hibernation may extend the service 

period of the detector up to an additional 

year. The service life will end once the 

detector reaches 24 months of actual 

operation. Hibernation provides a maximum 

service life of 36 months.

36-month Detector

3 years after activation, assuming 1 minute 

of alarm time per day.

Detection Range

H

2

S

 

0 to 100 ppm

CO

 

0 to 300 ppm

O

2

 

0 to 25% v/v O

2

 

SO

2

 

0 to 100 ppm

Factory Alarm Setpoints

 

Low Alarm      High Alarm

H

2

S

 

10 ppm 

    15 ppm

CO

 

35 ppm 

    200 ppm

O

2

 

19.5% v/v O

   23.5% v/v O

SO

2

 

5 ppm 

    10 ppm

Use Safety Suite Device Configurator via an 

IntelliDoX station to adjust alarm setpoints. 

For more information, refer to the operator 

manuals for Safety Suite Device Configurator 

software and the IntelliDoX automatic test 

and calibration station.

Ratings and Certifications

The BW Clip Real Time is in conformity with 

the following standards:
UL 913 8th Edition 

UL 60079-0:2013,  

UL 60079-11:2013
CSA C22.2 No. 157-92:2012,  

CSA C22.2 No. 60079-0:2011,  

CSA C22.2 No. 60079-11:2014
EN 60079-0:2012 +A11:2013, 

EN 60079-11:2012 

IEC 60079-0:2011 

IEC 60079-11:2011 

IEC 60079-26:2014 

ABNT NBR IEC 60079-0:2008 

ABNT NBR IEC 60079-11:2009

UL

Classified by UL to both US and Canadian 

Standards as intrinsically safe for Class 

I, Division 1, Group A, B, C, D and Class I, 

Zone 0, Group IIC.

ATEX

CE 0539 II 1G
Ex ia IIC T4 Ga IP66/67
DEMKO 14 ATEX 1356

CE

European Conformity

EU Declaration of Conformity

www.honeywellanalytics.com

IECEx

Ex ia IIC T4 Ga IP66/67
IECEx UL 14.0063

INMETRO

Ex ia IIC T4 Ga 
DNV 15.0024

KTL

15-KA4BO-0306

1

Specifications

Shelf Life

H

2

S

 

One (1) year before activation

CO

 

One (1) year before activation

O

2

 

Six (6) months before activation

SO

2

  

One (1) year before activation

Instrument Weight 

alligator clip included

92 grams (3.2 ounces)

Instrument Dimensions 

alligator clip included

41 x 50 x 87 millimeters 

(1.6 x 2.0 x 3.4 inches)

Operating Temperatures

H

2

S

 

-40°C to +50°C 

 

(-40 °F to +122°F)

CO

 

-30 °C to +50°C 

 

(-22 °F to +122°F)

O

2

 

-20 °C to +50°C 

 

(-4 °F to +122°F)

SO

2

 

-30 °C to +50°C 

 

(-22 °F to +122°F)

Intrinsic Safety Temperature Range

-40 °C to +50 °C (-40 °F to +122 °F)

Internal Vibrating

Operates to -10°C (+14°F)

Operating Humidity

5% to 95% relative humidity 

(non-condensing)

Audible Alarm

≈ 95 dB at 10 cm (3.9 in)

Visual Alarm

Flashing, wide-angled alarm lens with red 

LEDs plus alarm LCD readout

Display

Liquid crystal display (LCD)

Sensor Type

Electrochemical cells

Battery

Lithium, non-replaceable

Ingress Protection

IP 66/67

Alarm Setpoints

Instant low and instant high

Calibration

To ensure accurate calibration, Honeywell 

recommends using a premium-grade 

calibration gas approved by the National 

Institute of Standards and Technology (NIST).
Do not use a gas cylinder beyond its expiry date.

1

 

honeywell-analytics/products/bw-clip/docu-ments/

certificates/korea-certi ication_bw-clip-15ka-

44bo0306

 

Honeywell recommends you calibrate the 

detector at least once every 180 days depending 

on use and sensor exposure to poisons and 

contaminants. Calibrate only in a safe area that 

is free of hazardous gas.

Recommended Calibration Gas Mixtures

H

2

S

 

5 to 50 ppm balance N

2

CO

 

25 to 250 ppm balance N

2

SO

2

 

5 to 70 ppm balance N

2

O

2

 

5.0 % to 19.0 % v/v O

Perform the 

Zero

 procedure once every 

24 hours or when the 

Automatic Zero 

Reminder

 is displayed.

Intended Use

This product is classified for use in hazardous 

atmospheres that are not more than 21% v/v O

2

Recycling

This instrument contains a lithium battery. 

Do not mix with the solid waste stream. Spent 

batteries should be disposed of by a qualified 

recycler or hazardous materials handler.

a

 

CAUTION

Products may contain materials that are 

regulated for transportation under domestic 

and international dangerous goods regulations. 

Return product in compliance with appropriate 

dangerous goods regulations. Contact freight 

carrier for further instructions.

Limited Warranty and  

Limitation of Liability

Honeywell warrants this product to be free 

from defects in material and workmanship 

under normal use and service for a period of 

two or three years (depending upon detector), 

beginning on the date of activation. 24-month 

H

2

S and CO detectors are covered for up to 

an additional 12 months when hibernation is 

used, limited by a total of 24 months of detector 

operation. This Warranty is valid only if the 

detector is activated by the date on the package. 

This warranty extends only to the sale of new 

and unused products to the original buyer. 
The obligation of the warranty from Honeywell 

is limited, at the option of Honeywell, to refund 

of the purchase price, repair, or replacement 

of a defective product that is returned to a 

Honeywell authorized service center within the 

warranty period. In no event shall the liability 

of Honeywell liability hereunder exceed the 

purchase price actually paid by the buyer for 

the Product.

This warranty does not include: 
•  fuses, disposable batteries or the routine 

replacement of parts due to the normal wear 

and tear of the product arising from use;

•  any product which in the opinion of 

Honeywell, has been misused, altered, 

neglected or damaged by accident or 

abnormal conditions of operation, handling 

or use; or

•  any damage or defects attributable to repair 

of the product by any person other than 

an authorized dealer, or the installation of 

unapproved parts on the product. 

The obligations set forth in this warranty are 

conditional on: 
•  proper storage, installation, calibration, 

use, maintenance and compliance with the 
product manual instructions and any other 
applicable recommendations of Honeywell;

•  the buyer promptly notifying Honeywell of 

any defect and, if required, promptly making 
the product available for correction. No 
goods shall be returned to Honeywell until 
receipt by the buyer of shipping instructions 
from Honeywell; and

•  the right of Honeywell to require that the 

buyer provide proof of purchase such as the 
original invoice, bill of sale or packing slip 
to establish that the product is within the 
warranty period.

THE BUYER AGREES THAT THIS WARRANTY IS 

THE BUYER’S SOLE AND EXCLUSIVE REMEDY 

AND IS IN LIEU OF ALL OTHER WARRANTIES, 

EXPRESS OR IMPLIED, INCLUDING BUT 

NOT LIMITED TO ANY IMPLIED WARRANTY 

OF MERCHANTABILITY OR FITNESS FOR 

A PARTICULAR PURPOSE. HONEYWELL 

SHALL NOT BE LIABLE FOR ANY SPECIAL, 

INDIRECT, INCIDENTAL OR CONSEQUENTIAL 

DAMAGES OR LOSSES, INCLUDING LOSS OF 

DATA, WHETHER ARISING FROM BREACH OF 

WARRANTY OR BASED ON CONTRACT, TORT 

OR RELIANCE OR ANY OTHER THEORY.

¿¿¿’

Since some countries or states do not allow 

limitation of the term of an implied warranty, 

or exclusion or limitation of incidental or 

consequential damages, the limitations and 

exclusions of this warranty may not apply to 

every buyer. If any provision of this warranty 

is held invalid or unenforceable by a court of 

competent jurisdiction, such holding will not 

affect the validity or enforceability of any other 

provisions.

Summary of Contents for BW Clip Real Time

Page 1: ... BW Clip Factory Calibration Certificate This instrument was factory inspected tested and calibrated in accordance with the conditions and requirements of our registered Quality System Operating Standards and Sales Agreements MISE EN GARDE 1 Activez le détecteur avant la date d activation sur l emballage 2 Ce produit est un détecteur de gaz pas un dispositif de mesure 3 Veillez à assurer que la gr...

Page 2: ...s per second Over Limit and Under Limit Alarms If the gas concentration is beyond the sensor range OL over limit or OL under limit is displayed Audible 2 beeps per second Visual 2 flashes per second Vibration 2 vibrations per second Automatic Zero Reminder for O2 Models For O2 models only the Automatic Zero Reminder is displayed when more than 24 hours have elapsed since the last successful zero p...

Page 3: ... IntelliDoX station to enable the following optional settings for BW Clip Real Time detectors Non compliance After Gas Exposure When Non compliance after gas exposure is enabled the LEDs flash in an alternating pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints When the gas exposure is detected the top LEDs flash After 30 seconds the side LEDs flash After 30 sec...

Page 4: ...afe area that is free of hazardous gas Recommended Calibration Gas Mixtures H2 S 5 to 50 ppm balance N2 CO 25 to 250 ppm balance N2 SO2 5 to 70 ppm balance N2 O2 5 0 to 19 0 v v O2 Perform the Zero procedure once every 24 hours or when the Automatic Zero Reminder is displayed Intended Use This product is classified for use in hazardous atmospheres that are not more than 21 v v O2 Recycling This in...