background image

Zero the Sensor

Over time and through use, the sensor baseline at zero exposure may drift from 
the manufacturer’s baseline. For optimal performance of O2 models, Honeywell 
recommends that you zero the O2 sensor once every 24 hours or when the 

Automatic 

Zero Reminder

 is displayed. For all other models, Honeywell recommends that you zero 

the sensor periodically

Zero Procedure.

1.  Move to a normal atmosphere (20.9% v/v O

2

) that is free of 

hazardous gas.

2.  Press and hold the button until a 5 second countdown is 

displayed, and then continue to hold until the countdown is 
complete.

3.  When the countdown is complete, the 

Zero

 procedure begins 

and 

ZErO

 is displayed. 

4.  When 

Zero

 is successful, 

PASS

 and then 

CAL ?

 is displayed. 

Press the button to calibrate, or wait for the detector to enter 
normal operating mode after a 5 second countdown.

5.  If 

Zero

 is not successful, the non-compliance LED flashes. 

FAIL

 

and the non-compliance warning symbol are displayed. Press 
the button to acknowledge the result and return to normal 
operation. The non-compliance warning symbol is displayed and 
the non-compliance LED flashes. Perform the 

Zero

 procedure 

again. If the procedure fails again, contact Honeywell.

Bump Tests

When the detector is configured to display a bump test reminder, the bump test symbol 
is displayed when a bump test is due. For best results, bump test the detector via an 
IntelliDoX station. Bump tests performed via an IntelliDoX station are logged as bump 
tests. Manual bump tests are logged as unsafe gas concentrations.

Calibration

By default, BW Clip Real Time is configured to use the following calibration gas mixtures:

H

2

S: 20 ppm

balance N

2

CO: 100 ppm

balance N

2

SO

2

20 ppm

balance N

2

O

2

18.0 % v/v O

2

 

You can use Safety Suite Device Configurator software via an IntelliDoX station to change a 
detector’s default calibration gas mixtures. If the detector’s default calibration mixtures were 
changed, use the gas mixture values from Safety Suite Device Configurator to calibrate the 

detector.

1.  Move to a normal atmosphere (20.9% v/v O

2

) that is free of 

hazardous gas.

2.  Connect the calibration hose to the gas cylinder and to the 

intake inlet on the calibration cap.

3.  Press and hold the button until a 5 second countdown is 

displayed, and then continue to hold until de countdown is 
complete. The zero procedure begins and 

ZErO

 is displayed. 

When Zero is successful, 

PASS 

is displayed 

4.  When 

CAL ?

 is displayed, press the button to begin calibration.

5.  Put the calibration cap on the sensor grill.
6.  When 

GAS ?

 is displayed, apply calibration gas for two minutes 

at a recommended flow rate of 500 ml/min. When calibration is 
successful, 

PASS

 is displayed and the detector returns to normal 

operating mode.

 If calibration fails, non-compliance warning symbols are displayed. 

Press the pushbutton to acknowledge. Repeat calibration. If it fails again, contact 
Honeywell.

 

Event Logs

The detector stores the last 35 events that occurred, including peak readings, bump 
tests, calibrations and auto zeros. Each record contains:
•  detector serial number, sensor type and life-remaining;
•  total number of events that have occurred;
•  event type and duration;
•  alarm level(s) in ppm or %;
•  time elapsed since the alarm occurred in days/hours/minutes; 
•  duration of the alarm (minutes/seconds).
Use Safety Suite Device Configurator via an IntelliDoX station to transfer event logs from 
the detector to a computer. When logs are transferred via an IntelliDoX station, the most 
recent 35 events are transferred.

Optional Settings

Use Safety Suite Device Configurator via an IntelliDoX station to enable the following 
optional settings for BW Clip Real Time detectors.

Non-compliance After Gas Exposure

When 

Non-compliance after gas exposure

 is enabled, the LEDs flash in an alternating 

pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints. 
When the gas exposure is detected, the top LEDs flash. After 30 seconds, the side LEDs 
flash. After 30 seconds, the pattern repeats until more than 24 hours have passed since 
the last alarm, or when a successful bump test is performed via an IntelliDoX station.

Real-time Clock Display 

When 

real-time Clock Display

 is enabled, you can display the 

current time by pressing the button to 

Recall Events and Settings,

 

and then continuing to press the button until the time is displayed.
Time may be displayed in 12-hour or 24-hour clock format.

Hibernation

Hibernation

 is a factory installed feature that is available for 24-month H

2

S and CO 

detectors. When 

Hibernation

 is activated, the operating life countdown is suspended. 

Use 

Hibernation

 to deactivate the detector for 7 days or longer, up to a maximum of 12 

months. When 

Hibernation

 is activated, all detector safety functions are disabled.

Activate Hibernation via BW Clip Real Time 

Hibernation Case

1.  Have ready a BW Clip Hibernation Case. To purchase a Case, 
contact Honeywell or an authorized distributor.
2.  Move to a normal atmosphere (20.9% v/v O

2

) that is free of 

hazardous gas.
3.  Verify that the detector is in normal operating mode.
4.  Press the button to 

Recall Events and Settings

, and then 

continue to press the button until 

Hib

 is displayed.

5.  While 

Hib

 is displayed, press the button until a 5 second 

countdown is displayed. When the countdown is complete, the 
20-second 

Hibernation

 counter is displayed. 

6.  While the 

Hibernation

 counter is displayed, place the detector into a BW Clip 

Real Time Hibernation Case and then close the case firmly. The detector enters 

Hibernation

 mode. All detector safety functions are disabled.

7.  To reactivate the detector, open the case and remove the detector.

Activate Hibernation through Safety Suite Device 

Configurator via IntelliDoX

1.  Use Safety Suite Device Configurator to activate 

Hibernation on Insertion

 via an 

IntelliDoX station, and then insert the detector in the configured station. The detector 
event logs are retrieved, and the detector enters 

Hibernation

 mode. All safety 

functions are disabled, and the detector may be removed from the IntelliDoX station. 

2.  To reactivate the detector, press and hold the button until a 5 second countdown is 

displayed, and then continue to hold until the LCD and LEDs turn on and off. The 
detector performs a self-diagnostic test. When the test is successful, the type of gas 
detected and the remaining operating life are displayed. The detector is in normal 
operating mode. 

3.  For more information, refer to the operator manuals for the IntelliDoX station and 

Safety Suite Device Configurator software.

About this Publication

This publication is an operator guide for BW Clip Real Time portable single-gas detectors 

manufactured by Honeywell. It is intended for use with the following models:
•  24- and 36-month H

2

S and CO Detectors

•  24-month O

2

 detector

•  24-month SO

detector

Copyright, Notice, Trademarks

While this information is presented in good faith and believed to be accurate, Honeywell 

disclaims the implied warranties of merchantability and fitness for a particular purpose and 

makes no express warranties except as may be stated in its written agreement with and for its 

customers.

In no event is Honeywell liable to anyone for any indirect, special or consequential damages. The 

content in this document is subject to change.

BW Clip Real Time

2 OR 3 YEAR H

2

S, CO, O

2

, SO

2

 SINGLE-GAS DETECTOR

Summary of Contents for BW Clip Real Time

Page 1: ... BW Clip Factory Calibration Certificate This instrument was factory inspected tested and calibrated in accordance with the conditions and requirements of our registered Quality System Operating Standards and Sales Agreements MISE EN GARDE 1 Activez le détecteur avant la date d activation sur l emballage 2 Ce produit est un détecteur de gaz pas un dispositif de mesure 3 Veillez à assurer que la gr...

Page 2: ...s per second Over Limit and Under Limit Alarms If the gas concentration is beyond the sensor range OL over limit or OL under limit is displayed Audible 2 beeps per second Visual 2 flashes per second Vibration 2 vibrations per second Automatic Zero Reminder for O2 Models For O2 models only the Automatic Zero Reminder is displayed when more than 24 hours have elapsed since the last successful zero p...

Page 3: ... IntelliDoX station to enable the following optional settings for BW Clip Real Time detectors Non compliance After Gas Exposure When Non compliance after gas exposure is enabled the LEDs flash in an alternating pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints When the gas exposure is detected the top LEDs flash After 30 seconds the side LEDs flash After 30 sec...

Page 4: ...afe area that is free of hazardous gas Recommended Calibration Gas Mixtures H2 S 5 to 50 ppm balance N2 CO 25 to 250 ppm balance N2 SO2 5 to 70 ppm balance N2 O2 5 0 to 19 0 v v O2 Perform the Zero procedure once every 24 hours or when the Automatic Zero Reminder is displayed Intended Use This product is classified for use in hazardous atmospheres that are not more than 21 v v O2 Recycling This in...

Reviews: