background image

 

 

11 

 

9. TROUBLESHOOTING

 

 

Problem

 

Possible Cause 

Corrective Action 

No vibration 

 

settled intensity 
level zero(unit 

stand by) 

 

raise the set intensity level 

 

timer count down 

zero 

 

reset intensity level 

 

temperature 

higher than 

settled 

 

lower intensity level 

 

turn off unit 

 

non circuit loop 

 

finger touch the circuit 
polar  

 

stick a reference electrode 

 

unit broken 

 

call the dealer 

Intermittent 
vibration

 

 

temperature too 

high 

 

lower intensity level 

 

turn off unit 

 

bad circuit loop or 

too dry contract 

 

spread some gel or hydro-

cosmetic on the probe or 
skin 

 

contract circuit polar 

completely 

 

mixed wave mode 

 

settled ultrasonic mode 

unit off can 

no turn on 

 

unit inside 

humidity 

 

dry the unit more than 24 

hours 

 

DC connector 
loose contact or 

bad contact 

 

fixed connector completely  

 

replace new adaptor 

 

adaptor wire 

broken 

 

replace new adaptor 

 

unit broken 

 

contact the dealer 

 

NB: lf, after using this guide, the unit does not operate properly,the unit should be serviced. Please call 
your dealer for instruction.

 

 

 

 

 

 

 

 

Summary of Contents for HT906

Page 1: ...Manuale di utilizzo...

Page 2: ...INDICATIONS 5 4 INDICATOR AND CONTROLS 6 5 HOW TO USE THE CONDUCTIVE GEL 7 6 TECHNICAL SPECIFICATIONS 8 7 PREPARATION FOR USE 9 8 MAINTENANCE 10 9 TROUBLESHOOTING 11 10 WARRANTY 12 NB Before using the...

Page 3: ...fic areas of body ultrasonic energy at a frequency 1 M Hz and that is intended to generate deep heat within body tissues for the treatment INDICATIONS FOR USE The indications for use include the follo...

Page 4: ...cturer may cause harm This device is warranted for a period of 1 year from the date of purchase by the original buyer The device s shelf life is 3 years AC adaptor type 48 240 0300 CE Use timer max 15...

Page 5: ...ding use enough jelly to dissipate heat Do not submerge device and probe in water or solvent During use or after use if you feel unwell discontinue use and consult a physician If you have a preexistin...

Page 6: ...6 4 INDICATOR AND CONTROLS...

Page 7: ...ter or alcohol to clean it then spread the conductive gel on the skin or the device s vibration metal head Removal After use please use a tissue paper or soft and wet cloth to clean the skin and the v...

Page 8: ...c power 2 5W 20 max Temporal maximum average intensity 0 5W cm 2 20 max Timer 5 10 15 minutes three timer setting default levels 5 minutes Ultrasound Frequency 1 5 M Hz Pulsed Duty cycle L 12 5 M 33 H...

Page 9: ...button to choose the suitable timer setting 5 10 or 15 minutes There are 3 intensity levels to choose when the level is 15 press timer button to turn off the unit power indicator is off 5 Press mode...

Page 10: ...Conductive Gel Keep the conducive gel clean Be sure there is no dust oil or other foreign material on them If feel the gels dry too soon or can t attach well please spread more gel on the probe or sk...

Page 11: ...mperature too high lower intensity level turn off unit bad circuit loop or too dry contract spread some gel or hydro cosmetic on the probe or skin contract circuit polar completely mixed wave mode set...

Page 12: ...ch covers manufacturing flaws and defective materials on condition that the device is used properly and kept efficiently The warranty is limited to 12 months if the device is intended for professional...

Reviews: