background image

©

 Home Training Tools Ltd. 2005                                        

Page 5 of 8

                                  Visit us at www.homesciencetools.com

 

100x Objective 

suggested opening sizes for each power 
level:  

 

Objective Diaphragm 

Opening 

4x 

From fully closed to 1/8 open 

10x 

1/8 to 1/4 open  

40x 

1/4 to 1/2 open  

Note:  One method for finding the most suitable 
iris diaphragm opening is to remove the 
eyepiece, put your eye up to the eyepiece tube, 
and look at the bright circle of the aperture. 
Close the iris diaphragm until you see its 
outline in the aperture.  Adjust it so that its 
silhouette takes up 10-15% of the edge of the 
aperture.  Replace the eyepiece.  The contrast 
and resolution should be nicely balanced on 
your specimen.  If necessary, make further 
minor adjustments to the diaphragm while 
looking at the specimen.   

 
Setting the Field Diaphragm 

Even though your slide specimen is now in 

focus on the lowest power, you are not seeing 
the best image your microscope can give you.  
The following steps describe how to adjust your 
field diaphragm to set up the light path that will 
give you the sharpest image contrast.  Once it is 
set, you should be able to control your contrast 
and resolution using just the iris diaphragm.  Set 
your field diaphragm again only when you feel 
the contrast isn’t as sharp as it should be.   

1.

 

Focus a slide at the lowest power (40x).   

2.

 

Open the field diaphragm all the way by 
turning it clockwise.    

3.

 

Open the iris diaphragm all the way by 
rotating it clockwise.   

4.

 

Start to close the field diaphragm by turning 
it counter-clockwise.  At the same time, use 
the condenser knob to move the condenser 
up and down until the dark edge of the field 
diaphragm appears focused sharply. 

5.

 

Once the silhouette of the field diaphragm is 
sharply defined, center it using the 
condenser centering knobs (closing the field 
diaphragm further will help you center it).  

6.

 

Open the field diaphragm until its edge 
disappears beyond the outside of the field.   

7.

 

Now that your field diaphragm is set, you 
can turn up the light using the illumination 
intensity control. 

 
Changing Magnification  

1.

 

Use the stage controls to scan the slide 
(right to left and front to back) at low power 
to get an overview of the specimen.  Then 
center the part of the specimen you want to 
view at higher power.  

2.

 

Rotate the nosepiece to the 10x for 100x 
magnification (make sure it “clicks” into 
place). Refocus using the fine focus knob 
and view your specimen carefully. Adjust the 
iris diaphragm again until the image is most 
clear. Repeat with the 40x objective for 400x 
magnification. 

 

Using the 100x Oil Immersion Objective 

When using the 100x oil 

immersion objective, it is 
important to reduce light 
diffraction to enhance the image 
quality.  This is done by using 
immersion oil according to the 
following steps:  

1.

 

Lower the stage using 
the coarse focus knob.   

2.

 

Raise the Abbe condenser to the upright 
position by turning the condenser knob. 
Open the iris diaphragm by rotating it 
clockwise. 

3.

 

Place one drop of high quality immersion oil 
(our item MI-IMMOIL) on top of the 
coverslip. (In some cases, it may be 
beneficial to place a drop of oil between the 
condenser and the slide also, but this is 
usually not necessary for a good image.)   

4.

 

Move the 100x objective lens until it clicks 
into position, and then slowly move the 
stage up until the lens makes contact with 
the oil.   

5.

 

Continue focusing with the coarse knob until 
the color or blurred outline of the specimen 
appears.  Finish focusing with the fine focus 
knob.  Adjust the condenser and diaphragm 
until you see the specimen in clear, sharp 
contrast.   

6.

 

When you are done, clean all the oil off the 
lenses and slides using lens paper and lens 
cleaning solution (instructions are on page 2 
of this manual).  

 

 

Summary of Contents for MI-5000STD

Page 1: ...ate Home Microscope MI 5000DHD Ultimate Home Dual Head Microscope MI 5000BIN Ultimate Home Binocular Microscope 1 800 860 6272 www homesciencetools com Copyright 2005 by Home Training Tools Ltd All ri...

Page 2: ...Microscope Care Unpacking The Ultimate Home microscopes are shipped in a two part Styrofoam case Keep this case for storage transport and shipping It is perfect packing material should you ever need...

Page 3: ...ng loaded to prevent damage to the objective or slide The 100x objective is oil immersion this is described on page 5 6 Stage stop This is a screw with a lock nut located between the stage and the arm...

Page 4: ...work Plug the microscope s power cord into an outlet making sure that the excess cord is out of the way so no one can trip over it or pull it off of the table 2 Turn on your microscope s light source...

Page 5: ...centering knobs closing the field diaphragm further will help you center it 6 Open the field diaphragm until its edge disappears beyond the outside of the field 7 Now that your field diaphragm is set...

Page 6: ...e microscope using the inclined eyepiece and then use the diopter to adjust the camera image until it is sharp and clear on your monitor 2 A digital eyepiece package our item MI DIGICAM will include a...

Page 7: ...il you obtain a sharp image 4 Tighten the stop screw by turning it clockwise until it stops then turn it back a turn 5 Lock into position by tightening the locking nut Note The higher the stop screw i...

Page 8: ...Stage Acid and chemical resistant 130 x 140mm metal stage with stage clips and integral mechanical stage Mechanical Stage Low position coaxial x y controls provide precise 70mm movement on the x axis...

Reviews: