background image

©

 Home Training Tools Ltd. 2005                                        

Page 4 of 8

                                  Visit us at www.homesciencetools.com

 

light from the illuminator and makes the light 
change color slightly.  In some cases this 
may enhance the image of your specimen, 
though generally it is not necessary.   

11.

 

Coarse focus: 

The low-position large 

coarse focus knob is used to raise or lower 
the stage until the image is in focus.  The 
focus mechanism uses a slip clutch to 
prevent damage to the gears. 

12.

 

Fine focus: 

The smaller fine focus knob is 

coaxial with the coarse focus, which means 
that they turn on the same axis.  This knob 
allows more precise image resolving after 
the image has been brought into focus with 
the coarse focus knob.  

13.

 

Illuminator: 

The illuminator provides 

necessary light underneath the stage.  It 
contains a 20-watt halogen bulb. 

 

Instructions for changing the bulb are on 
page 7.   

14.

 

Field diaphragm:

  This diaphragm is 

located within the illuminator, creating a 
setup called Köhler illumination. The field 
diaphragm works with the iris diaphragm to 
produce an evenly illuminated field, a bright 
image without glare, and minimum heating 
of the specimen.   

15.

 

Illumination intensity control: 

  This 

control adjusts the intensity 
of the light produced by the 
illuminator.  It should be in a 
low position when turning the 
illuminator on or off.  Use the 
iris diaphragm to adjust the 
light for 

contrast;

 

use the illumination 

intensity control to adjust the light for 
brightness and intensity. 

Operating Procedure 

Now that you have an overview of what 

each component on your microscope is for, you 
can follow this step-by-step procedure to help 
you get started using it.   
 

Getting Started 

1.

 

Set your microscope on a tabletop or other 
flat sturdy surface where you will have 
plenty of room to work. Plug the 
microscope’s power cord into an outlet, 
making sure that the excess cord is out of 
the way so no one can trip over it or pull it 
off of the table.  

2.

 

Turn on your microscope's light source 
using the switch located at the rear of the 
base next to the cord. Open the iris 
diaphragm all the way by rotating it 
clockwise. 

3.

 

Rotate the nosepiece to the lowest-power 
(4x) objective.  You will hear a click when it 
is properly in place. Always start with the 
lowest power: it is easiest to scan a slide at 
a low setting, as you have a larger field of 
view.  

4.

 

Turn the coarse focus knob to move the 
stage down (away) from the objective lens 
as far as possible.  

5.

 

Move the small lever on top of the stage to 
open the curved slide clamp arm.     

6.

 

Carefully place a prepared slide squarely 
against the fixed arm and back edge of the 
mechanical stage.  Make sure the slide lies 
flat on the microscope stage.   

7.

 

Gently release the small lever allowing the 
slide clamp arm to securely hold the 
specimen slide in place. 

8.

 

Move the specimen under the objective lens 
by turning the stage control knobs. The 
upper knob is the front/back (y) adjustment 
control, which allows precise movement to 
the front and back of the stage. The lower 
knob is the right/left (x) adjustment control, 
which allows precise movement to the right 
and left of the stage.  Turn these controls 
without putting any pressure on the stage. 
This will allow you to scan a slide while 
maintaining fairly good focus. 

9.

 

Adjust the larger coarse focus knob until the 
specimen is in focus. Center the slide under 
the lens, if necessary, using the stage 
adjustment controls. Adjust the small fine 
focus knob until the specimen is clearly in 
focus. 

10.

 

Adjust the iris diaphragm to reduce light 
(turn it counter-clockwise) until the image 
has clear, sharp contrast. You will need less 
light on lower power and more light on 
higher power.  The following table gives 

Slide lever 

Slide clamp arm 

Fixed arm 

Summary of Contents for MI-5000STD

Page 1: ...ate Home Microscope MI 5000DHD Ultimate Home Dual Head Microscope MI 5000BIN Ultimate Home Binocular Microscope 1 800 860 6272 www homesciencetools com Copyright 2005 by Home Training Tools Ltd All ri...

Page 2: ...Microscope Care Unpacking The Ultimate Home microscopes are shipped in a two part Styrofoam case Keep this case for storage transport and shipping It is perfect packing material should you ever need...

Page 3: ...ng loaded to prevent damage to the objective or slide The 100x objective is oil immersion this is described on page 5 6 Stage stop This is a screw with a lock nut located between the stage and the arm...

Page 4: ...work Plug the microscope s power cord into an outlet making sure that the excess cord is out of the way so no one can trip over it or pull it off of the table 2 Turn on your microscope s light source...

Page 5: ...centering knobs closing the field diaphragm further will help you center it 6 Open the field diaphragm until its edge disappears beyond the outside of the field 7 Now that your field diaphragm is set...

Page 6: ...e microscope using the inclined eyepiece and then use the diopter to adjust the camera image until it is sharp and clear on your monitor 2 A digital eyepiece package our item MI DIGICAM will include a...

Page 7: ...il you obtain a sharp image 4 Tighten the stop screw by turning it clockwise until it stops then turn it back a turn 5 Lock into position by tightening the locking nut Note The higher the stop screw i...

Page 8: ...Stage Acid and chemical resistant 130 x 140mm metal stage with stage clips and integral mechanical stage Mechanical Stage Low position coaxial x y controls provide precise 70mm movement on the x axis...

Reviews: