TROUBLESHOOTING & MAINTENANCE 

Hygiene 

After using the product

Clean the device with a soft, damp cloth but make sure squeeze a cloth dry and to clean the foot pedal area.
Store the Electrode Gel pads on the plastic pad protectors provided.  

Storage

Keep the whole set of product clean and storage in a dust free and dry location under the condition 
Storage temperature and humidity

-10°C to 60°C, 10% to 95% RH

Operating temperature and humidity

-5°C to 50°C, 30% to 90% RH

Problem

Cause

Rectification

Device will not turn on. - Batteries inserted in wrong direction.

- The adapter does not plug well into device
properly.  

- Insert batteries in correct direction or check the
battery is in full power. 
- Check the connection of the adapter jack is in
well connected. And also the DC adapter with
well connecting to the main socket. 

Power turns off too
soon

- Gel pads not attached correctly to the skin.

- Attach Gel pads correctly to the skin

Power turns off while
using massager

- If you are using the batteries operation, then
the batteries weak/exhausted

- Fit 4 new identical 1.5V alkaline batteries type
AA

- The 30 minute timer is included as this is the
recommended maximum treatment time, it is
possible to simply restart the treatment if
required (for a second user for example).  

- Restart treatment or turn off the massager.

- If you are using the body massage, the
electrode gel pad may be broken

-Replace electrode gel pad

It is difficult to attach
Gel pad to the skin

- Transparent film not peeled off
- Gel pad applied immediately after washing
- Adhesive surface of Gel pad damaged
- The gel pads get dirty and lost their
adhesive/stickiness

- Peel off film on the adhesive surface of Gel pad
- Sufficiently dry Gel pad
- Replace Gel pad
- Replace Gel pad or clean the gel pad with a
small drop of water onto the sticky side of the
electrode pad and rub into the surface

Adhesive surface of
Gel pad is not sticky

- Use of Gel pad whilst perspiring
- Gel pad washed too long and/or too frequently
- Gel pads stored under high temperature, high
humidity, direct sunshine

Leave Gel pad in freezer for overnight

It is difficult to feel
stimulation

- You sole is too dry, not enough moisture 

- Your sole is not placing on the foot pedal
properly 
- Gel pads not attached correctly to the skin
- Gel pads overlap each other
- Electrode cord not connected correctly
- Applied intensity too weak

- Put some water to your sole to moisture your
sole
- Ensure both of your sole are placed on each
pedal properly. 
- Attach Gel pad firmly to the skin
- Reattach Long Life pads with no overlap
- Connect electrode cord correctly
- Increase the intensity by pressing the + button.

The skin turns red or
the skin feels irritated

- Adhesive surface of Gel pads dirty or dry

- Adhesive surface of Gel pads damaged

- Wash adhesive surface of Gel pads softly with
your fingertips for about 3 seconds under slow
running water
- Replace Gel pads

IB-CB200GB-1213-02_Layout 1  16/12/2013  12:02  Page 18

Summary of Contents for Circulator

Page 1: ...1 Instruction Manual Customer Service Number 44 0 1732 378 557 CB 200 GB IB CB200GB 1213 02_Layout 1 16 12 2013 12 02 Page 1 ...

Page 2: ...e battery on the remote control Connect the DC adapter to a suitable mains outlet and plug the small DC socket into the device Turn on the power the central display will light up orange and go off Remove your footwear and socks or stockings PLACE YOUR BARE FEET ONTO THE FOOT PADS YOUR RIGHT FOOT ON THE RIGHT FOOT PAD AND YOUR LEFT FOOT ON THE LEFT FOOT PAD BOTH FEET NEED TO BE ON THE DEVICE FOR IT...

Page 3: ... bare feet onto the footpads select the intensity setting and it automatically counts down from 30 minutes Q Am I too old to get any benefit from it A No The product is suitable for any adult age Q Can it really help me I m not very active and sit for most of the day A Yes As we sit the blood naturally pools to the lower legs due to gravity this is a natural action in the body If we do not take fr...

Page 4: ...ing to the environment and health Please dispose of batteries in designated collection points This marking indicates that this product should not be disposed with other household wastes throughout the EU To prevent possible harm to the environment or human health from uncontrolled waste disposal recycle it responsibly to promote the sustainable reuse of material resources To return your used devic...

Page 5: ...t is not functioning properly or you feel discomfort immediately stop using the unit If you feel any problems with your body or skin consult a doctor and follow his her instructions If you want to move the Electrode Pad to another region or your body during treatment be sure to turn off the power first If not you may receive a strong electrical shock Do not try to attach the Pads to any other pers...

Page 6: ...ssions Nevertheless special precautions need to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the unit and create a potentially unsafe situation CB 200 GB needs special precautions regarding EMC and needs to be installed and put into service according t...

Page 7: ...tes on safety Environment This unit is intended for home use only Effectiveness Massager relief of muscle pain stiffness and fatigue in the feet and lower legs Precautions for use Please read Notes on safety before using the unit Electronic Nerve Stimulation is a non invasive safe nerve stimulation intended to reduce pain The Circulator uses proven neuromuscular electrical stimulation therapy to s...

Page 8: ...ads 1 LCD Display Screen 2 Control Panel 3 Output for the feet Electrode Area for Left Foot 4 Output for the feet Electrode Area for Right Foot 5 Cable Connecting the Electrode Gel Pads and Device 6 Cable Connecting the Electrode Gel Pads and Device 7 Remote Control Receiver Sensor 8 Silver colour decoration plate 9 Adaptor Jack Top View Side View IB CB200GB 1213 02_Layout 1 16 12 2013 12 02 Page ...

Page 9: ...put intensity of body Available from 1 99 levels BODY Increase the output intensity of body Available from 1 99 levels SOLE Decrease the output intensity of Sole Available from 1 99 levels SOLE Increase the output intensity of sole Available from 1 99 levels TIME Decrease the operation time available from 1 60 minutes TIME Increase the operation time available from 1 60 minutes AUTO MODE Auto is t...

Page 10: ...repeating for 1 minute 9 Pulse rate 23 32Hz with 5 6 seconds and off time in 900mS A cycle repeating for 1 minute 10 Pulse rate 20 0Hz with 4 5 seconds and off time in 900mS A cycle repeating for 1 minute 11 Pulse rate 10Hz with 5 3 seconds and off time in 900mS A cycle repeating for 1 minute 12 Pulse rate 16 13Hz with 5 60 seconds and off time in 900mS A cycle repeating for 1 minute 13 Pulse rate...

Page 11: ...11 GB IB CB200GB 1213 02_Layout 1 16 12 2013 12 02 Page 11 ...

Page 12: ...12 GB IB CB200GB 1213 02_Layout 1 16 12 2013 12 02 Page 12 ...

Page 13: ...13 GB IB CB200GB 1213 02_Layout 1 16 12 2013 12 02 Page 13 ...

Page 14: ...ge from 1 60 minutes The timer will begin to count down from the time setting you select see Fig 3 To terminate the massage period user can turn off the unit anytime by pressing the on off Button once 5 If you are satisfied with the current massage programme you can lock the current massage programme by pressing the Auto Mode key The rest of the massage time will then only run the selected massage...

Page 15: ...ton of Body Or decrease intensity setting by pushing the button of Body The LCD will also show the level which you have selected see Fig 7 6 To terminate the massage period user can turn off the unit anytime by pressing the on off Button If you want to use with 2 gel pads only then you must connect 1 gel pad to jack A and 1 gel pad to Jack B ADDITIONAL ACCESSORIES Replacement gel pads For informat...

Page 16: ...ttery cover from the unit by removing the screw with the screwdriver Insert the new 2 pieces 1 5V size AAA batteries with the and marks correctly aligned The transmitter of the remote control is at the top of it in black colour Fig 10 for using the remote control please remember to point the transmitter to the receiver of the main unit which is located at between the gel pad jack of the main unit ...

Page 17: ...evice by using a soft cloth to clean the surface of the unit To clean the foot pedals use a soft damp soapy cloth but make sure squeeze a cloth dry and to clean the foot pedal area If the device is very dirty can use a soft damp soapy cloth but make sure squeeze a cloth dry to clean the unit Do not spill liquid on the device Do not immerse the device in water Do not clean with chemicals Store in a...

Page 18: ... gel pad may be broken Replace electrode gel pad It is difficult to attach Gel pad to the skin Transparent film not peeled off Gel pad applied immediately after washing Adhesive surface of Gel pad damaged The gel pads get dirty and lost their adhesive stickiness Peel off film on the adhesive surface of Gel pad Sufficiently dry Gel pad Replace Gel pad Replace Gel pad or clean the gel pad with a sma...

Page 19: ...tion 1 05 W Maximum Output Voltage U 54 8V during 1 kΩ load Maximum Output Current I 910 μA during 1 kΩ load Operating temperature and humidity 5 C to 50 C 30 to 90 RH Storage temperature and humidity 10 C to 60 C 10 to 95 RH Main unit dimensions 338 L x 324 W x 48 H mm Weight Approx 950 g Package Contents Quantity Parts 1 Circulator 1 AC DC Adapter 1 Remote Control 2 Cable Wire for Electrode Gel ...

Page 20: ... and manufacturer s declaration electromagnetic emissions The CIRCULATOR is intended for use in the electromagnetic environment specified below The customer or the user of the CIRCULATOR should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The CIRCULATOR uses RF energy only for its internal function Therefo...

Page 21: ...environment Surge IEC 61000 4 5 1 kV line s and neutral 1 kV line s and neutral Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 5 UT 95 dip in UT for 0 5 cycle Mains power quality should be that of a typical commercial or hospital env...

Page 22: ...ermined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures...

Page 23: ...to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is ...

Page 24: ...gned manufactured approved and or authorised or repair of products damaged by these modifications is not covered under this guarantee HoMedics shall not be responsible for any type of incidental consequential or special damages To obtain guarantee service on your product return the product post paid to your local service centre along with your dated sales receipt as proof of purchase Please refer ...

Reviews: