Guidance and manufacture’s declaration – electromagnetic immunity

The CIRCULATION PRO is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.

Immunity test

IEC 60601 test
level

Compliance level

Electromagnetic environment - guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 V/ms
150 kHz to 80
MHz

3 V/m
26 MHz to 2.5
GHz

10 V/m
26 MHz to 2.5
GHz

3 V/ms

3 V/m

10 V/m

Portable and mobile RF communications
equipment should be used no closer to
any part of the ELECTRO FLEX, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2 

P

d = 1,2 

P

80 MHz to 800 MHz

d = 2,3 

P

800 MHz to 2.5 GHz

where 

P

is the maximum output power

rating of the transmitter in watts (W)
according to the transmitter
manufacturer and 

d

is the recommended

separation distance in metres (m).

Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range
b.

Interference may occur in the vicinity of
equipment marked with the following
symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

a  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.

a

To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
CIRCULATION PRO is used exceeds the applicable RF compliance level above, the CIRCULATION
PRO should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the CIRCULATION PRO.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

13-0095 CB200-CA_IM_E_vE.qxd:Layout 1  7/2/13  2:05 PM  Page 21

Summary of Contents for CIRCULATION PRO

Page 1: ...na 2013 HoMedics LLC and its affiliated companies all rights reserved HoMedics is a registered trademark of HoMedics LLC and its affiliated companies All rights reserved 13 0095 CB200 CA_IM_E_vE qxd L...

Page 2: ...1 Instruction Manual CB 200 CA Circulation Pro 13 0095 CB200 CA_IM_E_vE qxd Layout 1 7 2 13 2 05 PM Page 1...

Page 3: ...ng the battery on the remote control Connect the DC adapter to a suitable power outlet and plug the small DC socket into the device Turn on the power The central display will light up orange and go of...

Page 4: ...ponents Notes on safety Danger This unit must not be used in combination with the following medical devices 1 Internally transplanted electronic medical devices e g pacemakers 2 Electronic life suppor...

Page 5: ...you feel any problems with your body or skin consult a doctor and follow his her instructions If you want to move the Electrode Pad to another region or your body during treatment be sure to turn off...

Page 6: ...use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the unit and create a pot...

Page 7: ...fety Environment This unit is intended for home use only Effectiveness Massager relieve of muscle pain stiffness and fatigue Precautions for use Please read Notes on safety before using the unit Elect...

Page 8: ...LCD Display Screen 2 Control Panel 3 Output for the feet Electrode Area for Left Foot 4 Output for the feet Electrode Area for Right Foot 5 Cable Connecting the Electrode Gel Pads and Device 6 Cable...

Page 9: ...tensity of body Available from 1 99 levels BODY Increase the output intensity of body Available from 1 99 levels SOLE Decrease the output intensity of Sole Available from 1 99 levels SOLE Increase the...

Page 10: ...cycle repeating for 1 minute 9 Pulse rate 23 32Hz with 5 6 seconds and off time in 900mS A cycle repeating for 1 minute 10 Pulse rate 20 0Hz with 4 5 seconds and off time in 900mS A cycle repeating fo...

Page 11: ...10 E 13 0095 CB200 CA_IM_E_vE qxd Layout 1 7 2 13 2 05 PM Page 10...

Page 12: ...11 E 13 0095 CB200 CA_IM_E_vE qxd Layout 1 7 2 13 2 05 PM Page 11...

Page 13: ...12 E 13 0095 CB200 CA_IM_E_vE qxd Layout 1 7 2 13 2 05 PM Page 12...

Page 14: ...om 1 60 minutes The timer will begin to count down from the time you selected see Fig 3 To terminate the massage period user can turn off the unit anytime by pressing the on off Button once 5 If you a...

Page 15: ...body 5 Gently increase intensity setting by pushing the button of Body Or decrease intensity setting by pushing the button of Body The LCD will also show the level which you have selected see Fig 7 6...

Page 16: ...Remove the battery cover from the unit by removing the screw with the screwdriver provided Insert 2 1 5V size AAA batteries with the and marks correctly aligned The transmitter of the remote control...

Page 17: ...oft cloth to clean the surface of the unit To clean the foot pedals use a soft damp soapy cloth but make sure to squeeze the cloth dry and clean the foot pedal area If the device is very dirty use a s...

Page 18: ...ifficult to attach Gel pad to the skin Transparent film not peeled off Gel pad applied immediately after washing Adhesive surface of Gel pad damaged The gel pads get dirty and lost their adhesive stic...

Page 19: ...U 54 8V during 1 k load Maximum Output Current I 910 A during 1 k load Operating temperature and humidity 5 C to 50 C 30 to 90 RH Storage temperature and humidity 10 C to 60 C 10 to 95 RH Main unit d...

Page 20: ...nufacturer s declaration electromagnetic emissions The CIRCULATION PRO is intended for use in the electromagnetic environment specified below The customer or the user of the CIRCULATION PRO should ass...

Page 21: ...ironment Surge IEC 61000 4 5 1 kV line s and neutral 1 kV line s and neutral the power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and volta...

Page 22: ...an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 M...

Page 23: ...ing to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For tr...

Page 24: ...RA from Customer Service and has sent the item along with its ORIGINAL receipt Shipping and taxes must be PRE PAID TO and HoMedics Group Canada by the PURCHASER This warranty gives you rights that var...

Reviews: