Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The [BPW-1005] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [BPW-1005] should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
BPW-1005, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2 √
—
P
80 MHz to 800 MHz
d = 2.3 √
—
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the [BPW-1005] is used exceeds
the applicable RF compliance level above, the [BPW-1005] should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
[BPW-1005].
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Summary of Contents for BPW-1005
Page 1: ...BPW 1005 EU Wrist Blood Pressure Monitor Instruction Manual...
Page 119: ...119 GB...
Page 120: ...120 GB...
Page 121: ...121 GB...
Page 122: ...122 GB...
Page 123: ...123 GB...
Page 124: ...124 GB...
Page 125: ...125 GB...
Page 126: ...126 GB...
Page 127: ...127 GB...
Page 128: ...128 GB...
Page 129: ...129 GB...
Page 130: ...130 GB...
Page 131: ...131 GB...