11. TROUBLESHOOTING (2)
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD shows low battery symbol
Low battery
Change the batteries
LCD shows “Er 0”
Pressure system is unstable before
measurement
Don’t move and try again.
LCD shows “Er 1”
Fail to detect systolic pressure
LCD shows “Er 2”
Fail to detect diastolic pressure
LCD shows “Er 3”
Pneumatic system blocked or cuff is
too tight during inflation
Apply the cuff correctly and
try again
LCD shows “Er 4”
Pneumatic system leakage or cuff is
too loose during inflation
LCD shows “Er 5”
Cuff pressure above 300mmHg
Measure again after five minutes.
If the monitor is still abnormal,
please contact the local distributor
or the factory.
LCD shows “Er 6”
More than 3 minutes with cuff
pressure above 15 mmHg
LCD shows “Er 7”
EEPROM accessing error
LCD shows “Er 8”
Device parameter checking error
LCD shows “Er A”
Pressure sensor parameter error
No response when you press button or
load battery.
Incorrect operation or strong
electromagnetic interference
Take out batteries for five minutes,
and then reinstall all batteries.