background image

©

 Home Training Tools Ltd. 2012                                        Page 6 of 8                                          Visit www.homesciencetools.com 

 

Locking nut 

Stop screw 

Tension collar

pressure on the stage. This will allow you to 
scan a slide while maintaining fairly good 
focus. 

 
Filters 

The MI-4100DXL model comes with a blue 

filter that can be placed in 
the filter holder located 
below the iris diaphragm. 
The filter absorbs some of 
the light from the 
illuminator and makes the 
light slightly blue. In some 
cases this may enhance the image of your 
specimen, though generally it is not necessary. 
Use the following procedure to insert the filter:  

1.

 

Locate the tiny filter holder knob underneath 
the iris diaphragm. Use this to swing the 
filter holder out from under the diaphragm.  

2.

 

Place the filter in the holder and swing back 
into place.  

3.

 

Adjust focus and diaphragm normally.  

 

Maintenance 

 

Adjusting the Stage Stop 

The stage stop is set at the factory to insure 

that the stage cannot come up far enough to hit 
the objective lenses. In normal circumstances 
you will not have to adjust this. However, if it 
falls out of adjustment or you are using a thinner 
slide that cannot be focused, follow these steps:    

1.

 

Loosen the round 
knurled locking 
nut by turning it 
counter-clockwise. 
You may need to 
use needle-nose 
pliers for this.  

2.

 

Loosen the stop 
screw.   

3.

 

Focus on a standard slide until you obtain a 
sharp image. 

4.

 

Tighten the stop screw by turning it 
clockwise until it stops, and then turn it back 
half a turn.  

5.

 

Lock into position by tightening the locking 
nut.  

6.

 

Note:  The higher the stop screw is, the 
higher the stage will rise.  

 
Changing the Bulb  

Your microscope bulb should last for 

years—approximately 8,000 hours. When it 
burns out, follow these steps for replacing it:   

1.

 

Obtain the correct 5-watt fluorescent 
replacement bulb (our item MI-BULB4. Note: 
some other 5-watt fluorescent bulbs may be 
too big).  

2.

 

Unplug your microscope from the power 
supply and allow it to cool before replacing 
the bulb.  

3.

 

Carefully lay the microscope on its side. 

4.

 

Using a screwdriver, 
remove the screw from the 
center of each rubber foot.  

5.

 

Remove the perforated 
metal plate and gently pull 
the bulb straight out from 
the socket.  

6.

 

Replace with a new bulb, 
then put the perforated 
metal plate back in place 
and replace the rubber feet.  

 
Adjusting Tension 

Coarse focus tension is pre-adjusted by the 

manufacturer, but if 
it falls out of 
adjustment, the 
stage will drift down 
under its own weight 
and the image will 
move out of focus. 
The tension adjustment collar is located 
between the microscope arm and the coarse 
focus knob on the right side (when the stage is 
facing you). To adjust the tension, follow these 
steps.  

1.

 

To tighten tension, turn the collar clockwise 
(you can get a better grip by putting a wide 
rubber band around the collar).  

2.

 

Tighten only enough to keep the stage from 
drifting downward. 

Summary of Contents for MI-4100STD

Page 1: ...Instruction Manual for MI 4100STD Home Microscope MI 4100DXL Home 1000x Microscope 1 800 860 6272 www homesciencetools com Copyright 2012 by Home Training Tools Ltd All rights reserved Home Microscope...

Page 2: ...Condenser 4 Iris Diaphragm 5 100x Oil Immersion Objective 5 Mechanical Stage 5 Filters 6 Maintenance 6 Adjusting the Stage Stop 6 Changing the Bulb 6 Adjusting Tension 6 Troubleshooting 7 Specificati...

Page 3: ...The 100x objective is oil immersion This is described on page 5 6 Stage The stage is the platform that supports the specimen slide below the objective lenses It moves up and down when you turn the foc...

Page 4: ...ough On MI 4100DXL open the iris diaphragm to the fully open position 3 Rotate the nosepiece to the lowest power 4x objective You will hear a click when it is properly in place Always start with the l...

Page 5: ...its lever counter clockwise Open the iris diaphragm by turning its lever clockwise 3 Place one drop of high quality immersion oil our item MI IMMOIL on top of the coverslip In some cases it may be ben...

Page 6: ...ide until you obtain a sharp image 4 Tighten the stop screw by turning it clockwise until it stops and then turn it back half a turn 5 Lock into position by tightening the locking nut 6 Note The highe...

Page 7: ...osepiece is not indexed properly Move revolving nosepiece until the objective lens clicks into position 2 The light is too bright Adjust the diaphragm Unable to focus slide 1 The slide coverslip is to...

Page 8: ...mm movement on the y axis Model MI 4100DXL only Condenser Fixed 0 65 N A condenser on model MI 4100STD Adjustable N A 1 25 Abbe condenser on model MI 4100DXL Diaphragm Calibrated 5 hole disc diaphragm...

Reviews: