6.4 TROUBLESHOOTING ETHERNET CONNECTIONS
Prior to proceeding below press CTRL+ALT+DEL to open Windows Task Manager.
Click the Processes tab. Ensure that there is only one instance of the application GUI
open. If there are more than one, end each Holzworth process, re-launch the GUI, and
attempt to establish a connection.
6.4.1 ETHERNET CONFIGURATION VIA USB & APPLICATION GUI
1.
If the synthesizer is not discovered by the application GUI there may be static
Ethernet settings that conflict with the current network configuration. USB
communication may be used to reset the synthesizer to DHCP or re-configure the static
network settings.
2.
Establish a USB connection with the synthesizer as shown in section
8.1.
3.
Launch the
Console
window using the button at the bottom right of the GUI. The
Console
can be used to send ASCII commands to change static network settings or
change from static mode to DHCP and vice versa.
4.
Refer to Appendix B for Ethernet configuration commands. Type commands into the
text field and then press Enter or click Send to send a command.
5.
Begin by querying with the :IP:STATUS? command. Change status and/or re-
configure the static network settings as necessary.
6.
Power cycle the HSX9000 if prompted. Any status change from DHCP to Static or
vice versa will require a power cycle.