24
System Operation
EN
CAUTION!
If a filled canister is removed from scale without activating the " Pause/Resume"
button, the message "Pause Pump for Canister Change" will appear, the pump
will stop immediately and the deficit display locked to insure an accurate deficit
count is maintained. Once the canister exchange is completed, the System is re-
started by pressing "Pause/Resume" button.
5.14
Total Volume Displayed
Total volume displayed
If a manual check of fluid deficit is desired, the total fluid volume can be obtained
by simultaneously pressing and holding both the up and down arrows (
(10)
&
(11)
in Figure 5-1, Front of Pump) underneath the Fluid Deficit Limit display (
(3)
in Figure 5-1, Front of Pump). The number in the Deficit Display is the total fluid
volume in ml. Once one or both of these buttons is released, the Deficit Display
will return to the Fluid Deficit value.
5.15
Turning System Off
Shut down
Press the
ON/OFF
switch to turn pump off. The displays and indicators are no lon-
ger illuminated.
WARNING!
Pressing the ON/OFF switch does not disconnect the system from the wall power
outlet. This requires pulling the power cord located in the rear of the system.