3.2.4 Affirm Table-top Stand
Caution:
The Affirm Table-top Stand is not applicable to mobile
applications.
To place the Affirm on the optional stand, refer to the following figure.
1.
Move the Affirm Biopsy Device Holder and needle guide to the highest upper
position as shown in
A
.
2.
Disconnect the Affirm cable from the Gantry.
3.
Remove the Affirm Guidance Module along with the Control Module, if present.
Caution:
To prevent damage or alignment problems with the
Needle Guidance Stage, be careful when you move the
Biopsy Guidance Module.
Caution:
The Affirm Biopsy Guidance Module weighs 15 pounds.
When you move it, be sure to have a secure grip on the
handles.
4.
Place the Affirm on the Table-top Stand as shown in
B
.
Figure 10:
Affirm Table-top Stand