Affirm Instructions for Use
Chapter 5
—
Troubleshooting
Audible Alerts
Part Number MAN-02866
41
Chapter 5
Troubleshooting
1.0 Audible
Alerts
Table
7:
Affirm
Audible
Alerts
Activity
Frequency
Duration
Repeats?
At
Power
Up:
3
250
ms
No
Any
Needle
position
within
the
safety
margin
limit:
1
50
ms
Yes
BGM
calibrate
or
configuration
operation,
Move
or
Jog:
•
If
the
operation
is
prevented
3
50
ms
No
•
If
the
operation
succeeds
1
100
ms
No
Start
of
each
Move,
including
Jog:
1
250
ms
No
2.0 Error
Messages
Table
8:
Affirm
Error
Messages
Error
Message
How
to
Correct
No
Device
Select
Select
a
biopsy
device.
No
Paddle
Install
a
biopsy
paddle.
User
is
not
logged
in
Log
in
at
the
Acquisition
Workstation.
Unit
is
unlatched
Put
both
lock
levers
into
the
locked
position
(see
the
figure
Installation
of
the
Biopsy
Guidance
Module
on
page
Unknown
Error
1.
Select
the
system
status
icon
in
the
taskbar
of
the
Selenia
Dimensions.
2.
Select
the
Clear
All
Faults
options.
3.
If
the
message
continues
to
display,
contact
Technical
Support.
Unrecoverable
Error
Contact
Technical
Support.
Software
must
be
reloaded.
Safety
Margin
Infringement
Move
the
biopsy
device
to
outside
of
the
safety
margin.
Press
the
Sound
button
to
mute
the
warning
sound.
Calibration
Required
Contact
Technical
Support.
Motion
Fault
1.
Select
the
system
status
icon
in
the
taskbar
of
the
Selenia
Dimensions.
2.
Select
the
Clear
All
Faults
options.
Another
message
that
is
related
to
the
cause
of
this
error
can
appear.
3.
If
the
Motion
Fault
message
continues
to
display,
contact
Technical
Support.