Affirm Instructions for Use
Chapter 2
—
How to Install or Remove the System
Installation of Components
Part Number MAN-02866
9
Chapter 2
How to Install or Remove the System
1.0
Installation of Components
1.1
How to Attach the Biopsy Guidance Module
You
can
install
the
Biopsy
Guidance
Module
with
the
Selenia
Dimensions
power
on
or
off.
Caution:
To
prevent
damage
or
alignment
problems
with
the
Needle
Guidance
Stage,
be
careful
when
you
move
the
Biopsy
Guidance
Module.
Caution:
The
Affirm
Biopsy
Guidance
Module
weighs
15
pounds.
When
you
move
it,
be
sure
to
have
a
secure
grip
on
the
handles.
Figure
7:
Installation
of
the
Biopsy
Guidance
Module
1.
Move
the
Compression
Device
down.
2.
Hold
the
Biopsy
Guidance
Module
by
both
handles.
3.
Slide
the
top
hooks
(item
2)
of
the
Biopsy
Guidance
Module
into
the
slots
marked
with
the
Affirm
icon
(item
2A)
on
the
front
of
the
Selenia
Dimensions
gantry.
Make
sure
the
top
and
bottom
hooks
attach
to
the
Selenia
Dimensions
gantry.
4.
Push
the
Lock
Levers
(item
3)
on
the
Biopsy
Guidance
Module
into
the
Up
position
to
lock
the
Module
against
the
Selenia
Dimensions.
5.
Align
the
red
dot
on
the
cable
(item
4)
from
the
Biopsy
Guidance
Module
with
the
red
dot
on
the
receptacle
on
the
Selenia
Dimensions
gantry.
Connect
the
cable
to
the
receptacle.