background image

function direction light , easy use in dark place .

2.2.13 HOLD

Push the ”HOLD” button to lock display value, push it again to
exit.

3. OPERATION
Warning

1)When measuring voltage ensure that instrument is not
connected or switched to a current or resistance Range, or to the
diode check, Always ensure that the correct terminals are used
for the type of measurement to be made.
2)Use extreme care when measuring voltage above 50V,
especially from sources where high Energy is existed.
3)Avoid making connections to “live” circuits whenever possible.
4)When making current measurements ensure that the circuit
not “live” before opening it in order to connect the test leads.
5)Before making resistance measurements or diode test, ensure
that the circuit under test is De--energized.
6)Always ensure that the correct function and range is selected.
If in doubt about the correct range to use, start whit the highest
and work downwards.
7)Extreme care should be taken when using the instrument to
conjunction with a current Transformer connected to the
terminals if an open circuit occurs.
8)Ensure that the test leads and probes are in good condition
with no damage to the insulation.
9)Take care not to exceed the over-lode limits as given in the
specification.
10) FUSE FOR REPLACEMENT MUST BE OF THE
CORRECT TYPE AND RATING.

11)

Before opening the case of the instrument to replace battery,

disconnect the test leads from any external circuit, set the
selector switch to

“OFF”

position.

3.1 DC Voltage measurement

1)Connect the black test lead to

“COM”

socket and red lead to

the

“VΩ”

socket.

2)

Set the selector switch to desired

DCV

position and connect

the probes across the source or load under measurement.

3.2 AC Voltage measurement

1) Connect the black test lead to

“COM”

socket and red lead to

the

“VΩ”

socket.

2)

Set the selector switch to desired

ACV

position and

connect the probes across the source or load under
measurement .

3.3 DC Current measurement

1) Connect the black test lead to

“COM”

socket and red lead to

the

“20A”

socket.

2)

Set the selector switch to desired current range position.

3)

Set the selector switch to desired

DCA

position and

connect the probes across the source or load under
measurement.

3.4 Resistance and continuity measurement

1) Connect the black test lead to

“COM”

socket and red lead to

the “

VΩ”

socket.

2)

Set the selector switch to desired resistance range position

or “

” position.

3) Connect the probes across circuit to be tested.
Caution: Ensure that the circuit to be tested is “dead”.
Max. input over-load: 250V RMS And<10sec.

3.5 Diode Test

1)

Connect the black test lead to

“COM”

socket and red test

leads to the

“VΩ”

socket.

2) Set the selector switch to

“ ”

position .

3)

Connect the black and red test probe to cathode

(-)

and

anode

(+)

ends of diode to be tested respectively.

3.6 Temperature measurement

Set the selector switch to the

‘TEMP

/

range and connect

the K type thermocouple’s black test lead to

“TEMP-”

socket and

red test lead to the

“T+”

socket.

A. The temperature function shows the random

number at ordinary times, must insert the thermocouple

in temperature test hole while examining temperature. In

order to guarantee the accuracy of the measurement,

need to close Light switch while measuring temperature.

B. This meter in closure WRNM-010 type contact

thermocouple limit temperature is 250

(300

shortly ) ;

C.

Please don't change the thermocouple at will ,

otherwise we can't guarantee to measure accuracy ;

D.

Please

don’t

importing

the

voltage

in

the

temperature function.

3.7 RPM (TACH ) measurement

1). Connect the black test lead to

“COM”

socket and red test

leads to the

“VΩ”

socket.

2). Connect the red test probe to “BREAKER POINTS ” or

”terminal of IGNITION CIOL ,and connect the black probe to

“GROUND ”or “ - ” terminal of battery .
3). Set the selector switch to the

“TACH

” position for 3cyl,4cyl ,

5cyl , 6cyl , 8cyl , according to the engine cylinders .For any
TACH signal with a given frequency , in convention
ignition system

RPM=120×freq÷cyl . in DIS ignition system :

RPM=120×freq÷2 .
4). Start the engine and read engine RPM .

3.8 DWELL ANGLE measurement

1). Connect the black test lead to

“COM”

socket and red test

lead to the

“VΩ”

socket .

2). Connect the red test probe to “BREAKER POINTS ” or
“ ”terminal of IGNITION CIOL ,and connect the black probe to
“GROUND ”or “-” terminal of battery .
3).

Set the selector switch to the

“DWELL”

position for 3cyl,

4cyl , 5cyl , 6cyl , 8cyl , according to the engine cylinders .
4). Start engine the display will show the DWELL ANGLE of
points.

4.

ACARE AND MAINTENANCE

4.1 CARING FOR YOUR MULTIMETER

Your digital multimeter is an example of superior design and

craftsmanship. The following Suggestions will help you care for
the milimeter so you can enjoy it for years.
1). Keep the multimeter dry. If it gets wet , wipe it dry immediately.
Liquids can contain minerals that Can corrode electronic circuits
2).

Use and store the multimeter only in normal temperature

environments, Temperature extremes Can shorten the life of
electronic devices, damage the batteries, and distort

or melt

plastic parts.
3). Handle the multimeter gently and carefully. Dropping it can
damage the circuit boards and case and can cause fthe
multimeter to work improperly although the holster can provide
enough protection.
4). Keep the multimeter away from dust and dirt, which can
cause premature wear of ports.
5). Wipe the multimeter with a damp cloth occasionally to keep it
looking new. Do not use harsh Chemicals, cleaning solvents, or
strong detergents to clean the multimeter.
6). Use only fresh batteries of the required size and type .always
remove old or weak batteries. They Can leak chemicals that
destroy electronic circuits.

4.2 MAINTENANCE
9-volt battery replacement

a)

Ensure the instrument is not connected to any external

circuit. snap fastener “OFF” Position and remove the test leads
from the terminals.
b)

Remove the screw on the bottom case and lift the battery

case. Remove the spent battery and Replace it with a battery of
the same type.
c)

Please take out the battery when not using for a

long time.

Above picture and content just for your reference.

Please be subject to the actual products if anything

different or updated. Please pardon for not

informing in advance.

Reviews: