Calibration instructions
geodyna 6300
Page 8 of 8
Date 2 Oct 2001
Storing calibration data with code C90
Service and test functions become accessible by simultaneously pressing and
holding the C and the balancing mode keys and by rotating the main shaft.
The following reading appears:
Press the precision key and rotate
the main shaft to set 1.
The following reading appears:
Press the C key to complete calibration.
The following reading appears:
Press the STOP key to quit the service mode.
C 90. 0
C 90. 1
C - -