2. P-Unit
The device can show the pressure in the
pressure units hectopascal (hPa), millibar
(mbar), and und centimeters of water (cmH
2
O).
For changing the pressure unit go to the Clinic
menu (hold down the
button for approx. 1
second). Then select »P-unit« by pressing the
buttons
and
. Press the
button. Each
time you press the button another pressure unit
will appear. Leave the menu once the desired
pressure unit is selected. The device will then
save the new unit.
PLEASE NOTE!
When changing the pressure unit from hPa or
mbar to cmH
2
O the pressure will slightly
change.
Example: 8 hPa = mbar are set.
At 8 cmH
2
O the pressure is 7.8456 hPa/mbar.
1 cmH
2
O = 0.9807 hPa.
3. Turbine
This menu item shows the total operating time
of the turbine, the filter, and the therapy time.
3.1 Turbine
The counter indicates the total turbine running
time.
3.2 Filter
The counter indicates the operating hours of
the filter since the last reset.
3.3 Therapy
The counter indicates the operating time with
the therapy pressure applied.
3.4 Standby
This counter indicates the time the device was
in the standby mode.
To display the counter go to the Clinic menu
(hold down the
button for approx. 1 second).
Then select »Turbine« by pressing the buttons
and
. Press the
button.
Each time you press the button now you see
the next counter.
30