Technical data
71
TECHNICAL DATA
VECTOR et
VECTOR et with AquaDROP et
Dimensions
230 x 212 x 107 (W x D x H) mm
230 x 342 x 125 (W x D x H) mm
Weight
approx. 2 kg
approx. 2.35 kg (without water)
Power supply
100 V ~ … 240 V ~, 50 ... 60 Hz, 24 V DC / 2.1 A
Operating time
under battery power
up to 8 years
Pressure range
4…20 hPa (mbar); BILEVEL ST30: 4…30 hPa (mbar)
Max. limit pressure
in the event of a fault
30 hPa (CPAP - APAP)
40 hPa (BILEVEL - BILEVEL ST20 und ST30)
Power consumption
DC power
Mains power
DC power
Mains power
Standby
< 5 W
< 5 W
---
---
Operation at 12 hPa
8 W
10 W
28 W
32 W
Short term pressure
variation
4 hPa
10.5 hPa
17 hPa 23.5 hPa
30 hPa
4 hPa
10.5 hPa 17 hPa
23.5 hPa
30 hPa
10 bpm
0.15 hPa
0.16 hPa (at heating level 3)
15 bpm
0.30 hPa
0.30 hPa (at heating level 3)
20 bpm
0.46 hPa
0.49 hPa (at heating level 3)
Long term pressure
variation
± 0.3 hPa
Pressure reading
accuracy
0.5 hPa
Average sound pres-
sure level (
operating
at 1 m distance)
25 dB(A) at 10 hPa
(equivalent to a sound power level of
33 dB[A])
Air
fl
ow rate
4 hPa
> 180 l/min
7.5 hPa
> 170 l/min
15 hPa
> 160 l/min
22.5 hPa
> 150 l/min
30 hPa
> 140 l/min
Operating
temperature
+ 5 °C ... + 40 °C
Storage temperature
- 20 °C ... + 70 °C