humidifier
AquaTREND III-H5
AquaTREND III-H5
should be thoroughly flushed once
with filtered water before first use. In this regard, read
the section “Cleaning” on page 9.
AquaTREND III-H5
must not be fitted to therapy
device during filling.
Open the humidifier by turning the cap anticlockwise
and the reservoir in opposite direction.
Fill
AquaTREND III-H5
only with distilled water or boi-
led water with a low lime content. This water must be
cold and contain no additives. The maximum filling level
must not be exceeded.
Close the humidifier by puting the cap on the reservoir
and turning clockwise until limit stop.
A cleaned and freshly filled humidifier must no longer be
placed on its side or turned upside down. Take the
humidifier over to your
TREND
device and fit it.
In so doing, ensure the following:
n
Press the release button on top of the therapy
device
n
Feed the contact pins and air inlet to the therapy
device without turning or tilting the humidifier.
n
The rear panel of the humidifier must come flush
with the
TREND
without any gap.
n
The release button must be heard to
engage.
On
AquaTREND III-H5
the heating level is selectable
in 6 stages from 0 to 5. By touching the touch pad the
heating level will increase at a time for one stage until 5
and then begin with 0.
The adjusted heating level is shown on the LED. Heating
of the water occurs only when the fan is operating. The
heating is switched off in heating stage “0”.
Caution!
Starting the blower with the humidifier full and without
the therapy hose connected can cause the air flow to
pick up drops of water. Therefore start the humidifier
only when the therapy hose is properly inserted.