5

Start the run.

 Press the START/STOP button. The 

green light above the START/STOP button indicates 
that voltage and current are being applied to the 
output jacks. The green light above the CONSTANT 
button indicates which parameter (V or A) is limiting, 
and constant. The LED display shows the numerical 
value of the actual V or A output, or the time (press 
the SELECT button to toggle between the values). 

6

In the case of a continuous run, press START/STOP to  
terminate the run. 

The LED display reads “OFF”.

A timed run automatically stops with a warning beep. 
The word “OFF”, and the conditions at the end of the 
run are flashed on the display so that the user can 
record the end conditions. 

Pushing the SELECT button will return the power 
supply to the last run settings.

Pressing the CONSTANT button will reset the power 
supply.  

Example: Set up conditions for running 2 minigels, 
each at a constant current of 30 mA.

1   Press the CONSTANT button to position the 

indica tor light under “A”.

2   Press the UP and DOWN buttons to set the LED 

display value 60. (2 minigels × 30 mA each).  

3   Press the SELECT button to position the indicator 

light under “V”. Observe that the volt limit has 
been automatically set to the maximum value of 
300 V.

 

 

Setting the voltage limit to the maximum setting 
ensures that the entire separation is run at 60 mA 
for as long as possible. Should the buffer or gel 
lose too much conductivity during the run, the 
opera tional 300 V or 90 W limits may be reached, 
at which point “crossover” automatically occurs 
and V becomes the limiting variable. The current 
would slowly drop below 60 mA at this point.

 

 p10

Note: 

The power supply slowly 

increases to the set value when 
starting a run. 

The constant parameter can be 
adjusted during a run. 

Summary of Contents for PS300B

Page 1: ...user manual um PS300B IM Rev C0 07 12 PS300B Power Supply...

Page 2: ...formation viii Power supply function and description 1 Unpacking 3 Specifications 4 Parameter overview 5 Power supply operation 7 Connections 8 Operation 9 Power Failure Detection PFD 12 Troubleshooti...

Page 3: ...b t a smrtiacej ok p e mus b t vykon v na v jeho provoz Tento n stroj je ur en v souladu s EN61010 1 2001 elektrick bezpe nostn normy P esto e by m ly b t pou ity pouze dn vy kolen oper tor st cel tot...

Page 4: ...te staan voldoende afkoelen verzeker dat de luchtopeningen van het instrument niet bedekt zijn T rke Tietoa Finnish Jos t t varusteita k ytet n tavassa ei m ritetty Hoeferille Inc suojelu ehk isty va...

Page 5: ...den Nur unbesch digte elektrische Leitung und Ausr s tung spezifisch f r die Stromspannungen verwenden die Sie verwenden werden Die ganze mit der Hoch spannung verbundene Ausr stung sollte in berein s...

Page 6: ...nstrument ten jest przeznaczony do u ytku w labora toriach kryty tylko Tylko akcesori w i cz ci zatwierdzone lub dostarczone przez Hoefer Inc mog by wykorzystane do eksp loatacji utrzymania i obs ugi...

Page 7: ...ida de red Utiliza alambre y equipo el ctricos s lo ilesos espec ficos para los voltajes que usted utilizar Todo equipo conectado al voltaje alto debe ser de acuerdo con EN61010 1 2001 Mantiene el ins...

Page 8: ...Kontakt mit einem autorisierten Beauftragten des Herstellers auf um Informationen hinsichtlich der Entsorgung Ihres Ger tes zu erhalten Questo simbolo indica che i rifiuti derivanti da apparecchiatur...

Page 9: ...ct CE certification This product meets the requirements of applicable CE directives A copy of the corresponding Declaration of Conformity is available on request The CE symbol and corresponding declar...

Page 10: ...in combi nation with a SELECT button Four pairs of output jacks accept 4 mm plugs which operate in parallel The output jacks are color coded recessed for safety and accept 4 mm shielded banana plugs t...

Page 11: ...p2 Fig 1 Front view of PS300B Power Supply 3 digit display UP DOWN arrows SELECT display highlighted parameter CONSTANT V or A toggle START STOP polarity Output jacks polarity...

Page 12: ...fer Inc sales office Inspect all components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing mate...

Page 13: ...l Output connector 4 mm plugs Timer 1 999 min 1 min increment Environmental operating Indoor use 4 40 C conditions Relative humidity 0 80 noncondensing Altitude 2000 m Pollution degree 2 Installation...

Page 14: ...oes not exceed the power supply rating Crossover occurs when the second parameter limit is reached as system resistance changes Fig 3 At this point the second parameter becomes constant and the previo...

Page 15: ...wer When this occurs both the constant V and the constant mA lights will be lit at the same time p6 current Voltage limit Crossover from constant A to constant V Current limit resistance increasing ov...

Page 16: ...ng or discon necting the voltage leads 2 Keep fingers away from metal components when connecting or disconnecting the voltage leads p7 WARNING Use only one hand when making or breaking a connection to...

Page 17: ...e power supply so that the fan on the back panel and the vents on the bottom are not obstructed 2 Connect the HV power leads Connect the apparatus to the power supply by plugging the lead connectors i...

Page 18: ...s to set it to the desired value The power supply can then operate in Crossover mode as explained on page 5 4 Set the timer optional Press the SELECT button to move the green indicator light to under...

Page 19: ...sing the CONSTANT button will reset the power supply Example Set up conditions for running 2 minigels each at a constant current of 30 mA 1 Press the CONSTANT button to position the indica tor light u...

Page 20: ...ELECT button to position the indicator light under the Clock symbol If desired use the UP button to set the timer to a desired duration If the run is continuous set the timer value to 0 5 Press the ST...

Page 21: ...The time shown at the end of the run is the time the gel ran since power was restored not the total time of the run The default settings for Power Failure Detection are Run Time Setting PFD Code Time...

Page 22: ...used The output current may have fallen below the 4 mA limit of the power supply Press any button to clear the error code E4 Power Failure during a continuous run no timer being used Restart the run i...

Page 23: ...l Use a screwdriver to pry out the fuse compartment There are two fuses in the compartment Check and replace any blown fuses with a fuse of the same size and ratings FUSES 2 250 V 3 15 A 5 20 mm fast...

Page 24: ...ll Hoefer Inc for advice before using Technical service and repair Hoefer Inc offers complete technical support for all of our products If you have any ques tions about how to use this product or woul...

Page 25: ...lliston MA 01746 Toll Free 1 800 227 4750 Phone 1 508 893 8999 Fax 1 508 893 0176 E mail support hoeferinc com Web www hoeferinc com Hoefer is a registered trademark of Hoefer Inc 2012 Hoefer Inc All...

Reviews: