background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 8 

www.onsetcomp.com 

Only the device used to set the password can then connect to 
the logger without entering a password; all other devices are 
required to enter the password. For example, if you set the 
password for the logger with your tablet and then try to 
connect to the logger later with your phone, you must enter 
the password on the phone but not with your tablet. Similarly, 
if others attempt to connect to the logger with different 
devices, they are also required to enter the password. To reset 
a password, press the switch on the logger for 10 seconds or 
connect to the logger and tap Manage Password and tap Reset. 

Downloading Data From the Logger 

 To download data from the logger: 

1.

 

Tap Devices.  

2.

 

Tap Download Data. The logger downloads data to the 
phone, tablet, or computer.  

3.

 

Once the download is complete, tap Done to return to the 
previous page or tap Export and Share to save the file in the 
specified format. 

4.

 

When the export file has successfully been created, tap 
Done to return to the previous page or tap Share to use 
your device’s usual modes of sharing. 

You can also upload data automatically to HOBOlink, Onset’s 
web-based software, using the app or the MX gateway. For 
details, see the HOBOconnect User Guide and see the HOBOlink 
help for details on working with data in HOBOlink. 

Logger Events 

The logger records the following events to track logger 
operation and status. You can view events in exported files or 
plot events in the app. 

To plot events, tap Data and select a file to open.  

Tap   (if applicable) and then tap 

. Select the events you 

want to plot and tap OK. 

Internal Event Name 

Definition 

Host Connect 

The logger was connected to the mobile 
device. 

Started 

The logger started logging. 

Stopped 

The logger stopped logging. 

Calibration End 

pH calibration ended while the logger was 
logging. 

Calibration Cancel 

pH calibration was canceled while the 
logger was logging. 

Calibration Values 

The slope and offset values that were set 
during calibration while the logger was 
logging. 

Calibration Buffers 

The buffers used during calibration while 
the logger was logging. 

Alarm 
Tripped/Cleared 

An alarm has occurred because the reading 
was outside the alarm limits or back within 
range. 

Note:

 Although the reading may 

have returned to a normal range during 
logging, an alarm cleared event will not be 
logged if the logger was set up to maintain 
alarms until reconfigured. 

Internal Event Name 

Definition 

New Interval 

The logger has switched to logging at the 
burst logging rate or back to the normal 
rate. 

Water Detect 

The logger has been placed in water or 
removed from water. 

Low Power 

The battery level dropped below 0.8 V. 

Safe Shutdown 

The battery level dropped below a safe 
operating voltage and the logger 
performed a safe shutdown with all data 
stored in memory. 

Deploying and Mounting the Logger 

Important:

 

The pH sensor must be kept in storage solution 

when not being calibrated or deployed in water

Follow these guidelines when deploying the logger: 

 

If the logger is in storage solution, remove the clear 
storage cap from the logger and replace it with a closure 
cap as shown in the next diagram before deploying it. 

 

 

If your water sample is prone to biofouling, insert the 
optional copper anti-biofouling guard over the pH sensor 
as shown. Gently twist the copper guard so that the holes 
line up with the holes in the sensor end cap and pH 
electrode. Do not use the copper guard when the pH 
sensor is being calibrated or in storage solution, or if the 
sensor is placed in a very small container with still water 
or in water where copper can affect the environment. 

 

Line up holes in copper guard with holes 
in sensor end cap and pH electrode 

Optional copper guard 
installed to protect 
against biofouling 

Make sure the closure 
cap is installed before 
deploying the logger 

Summary of Contents for MX2501

Page 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Page 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Page 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Page 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Page 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Page 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Page 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Page 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Page 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Page 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Page 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Page 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Reviews: