Hobo MX2501 Manual Download Page 10

HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 10 

www.onsetcomp.com 

3.

 

Use a foam swab dipped in water to clean the internal O-
rings. Use the foam swab to spread a small dot of silicone-
based grease along both the O-rings inside the sensor end 
cap. Do not use a cotton swab as the fibers on the swab can 
interfere with the operation of the electrode. 

 

4.

 

Remove the replacement pH electrode from the shipping 
bottle and wipe the outer plastic body dry with a soft paper 
or cloth towel. 

Never touch the glass bulb inside the clear 

plastic end of the probe.

 

5.

 

Remembering not to touch the glass bulb, push the pH 
electrode 

gently and slowly

 into the sensor end cap. You 

will feel the pH electrode pause as it passes through two 
internal O-rings inside the sensor end cap (shown with step 
3). Make sure the internal O-rings are still in place as the 
probe is pushed into the sensor end cap. 

Note:

 If done too 

quickly, this process can unseat the internal O-rings. 

Gently twist the electrode so that the holes in the clear 
electrode body line up with the holes in the end cap. 

 

6.

 

Continue to gently push the electrode into place until the 
clear plastic is fully seated in the end cap as shown. 

 

7.

 

Push the sensor end cap back into the logger, making sure 
the small tab in the end cap fits in the small groove in the 
logger housing. Use the large tab on the end cap and the 
arrow on the logger housing to help with proper alignment. 
Do not twist or turn the end cap as you push it straight into 
position. 

 

 

8.

 

If you will be calibrating the logger immediately:

 screw the 

closure cap back on the logger and follow the instructions in 

Calibrating the Logger

If you will not be calibrating or deploying the logger 
immediately:

 

a.

 

Pour storage solution into the clear storage cap to the 
marked fill line. 

 

b.

 

Carefully screw the cap with storage solution onto the 
logger. 

 

Important:

 Keep the sensor in storage solution unless you are 

calibrating the logger or deploying it in water. Only use storage 
solution in the cap; do not use water. Do not use the copper 
guard while the sensor is in the cap. 

Battery Information 

The logger requires one user-replaceable AA 1.5 V alkaline or 
lithium battery. With a logging interval of 1 minute and at 25°C 
(77°F)

battery life is 1 year with Bluetooth Always On selected 

and 3 years with Bluetooth Always Off selected. Expected battery 
life varies based on the ambient temperature where the logger is 
deployed, the logging interval, the frequency of connections, 
downloads, and paging, and the use of burst mode or statistics 
logging. Deployments in extremely cold or hot temperatures or a 
logging interval faster than 1 minute can impact battery life. 
Estimates are not guaranteed due to uncertainties in initial 
battery conditions and operating environment.

 

pH electrode installed 

Use a foam swab dipped in 
water to clean the internal 
O-rings; use a foam swab 
to spread a dot of silicone-
based grease on the 
internal O-rings 

Push the end cap 
back into logger, 
aligning tabs with 
arrow 

Sensor end cap 
installed with 
correct alignment 

Large tab 

Small tab 

Small groove 

Gently and slowly

 push pH 

electrode into end cap and 
twist the electrode so the 
holes in the electrode align 
with the holes in the end cap 

Pour storage 
solution into the 
cap to the fill line 

Screw on the 
filled cap 

Summary of Contents for MX2501

Page 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Page 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Page 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Page 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Page 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Page 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Page 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Page 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Page 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Page 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Page 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Page 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Reviews: