background image

HOBO MX2300 Series Data Logger Manual 

1-800-LOGGERS 

www.onsetcomp.com 

 

6.

 

Enable Low Limit if you want burst logging to occur when 

the sensor reading falls below a specific reading. Drag the 

slider to the reading that will trigger burst logging or tap the 

value field and type a specific reading. In the example, the 

logger will switch to burst logging when the temperature 

falls below 32°F. 

7.

 

Tap Done and repeat steps 4–7 for the other sensor if 

desired. 

8.

 

Tap Burst Logging Interval and select an interval faster than 

the logging interval. Keep in mind that the more frequent 

the burst logging rate, the greater the impact on battery life 

and the shorter the logging duration. Tap Done. 

9.

 

Tap Done to exit the Logging Mode screen. 

10.

 

Tap Start in the Configure screen to load the burst settings 

onto the logger if you are ready to start. 

Notes: 

 

Burst limits are checked <placeholder>.

 

 

If high and/or low limits have been configured for more 

than one sensor, then burst logging will begin when any 

high or low condition goes out of range. Burst logging will 

not end until all conditions on all sensors are back within 

normal range. 

 

The actual values for the burst logging limits are set to 

the closest value supported by the logger. For example, 

the closest value to 85°F that the logger can record is 

84.990°F and the closest value to 32°F is 32.043°F. 

 

Burst logging mode can begin or end when the sensor 

reading is within the logger specifications of 0.04°C 

resolution. This means the value that triggers burst 

logging may differ slightly than the value entered. For 

example, if the high limit for a temperature alarm is set 

to 75.999°F, burst logging can start when the sensor 

reading is 75.994°F (which is within the 0.04°C 

resolution). 

 

Once the high or low condition clears, the logging 

interval time will be calculated using the last recorded 

data point in burst logging mode, not the last data point 

recorded in “normal mode.” For example, let’s assume 

the logger has a 10-minute logging interval and logged a 

data point at 9:05. Then, the high limit was surpassed 

and burst logging began at 9:06. Burst logging then 

continued until 9:12 when the sensor reading fell back 

below the high limit. Now back in normal mode, the next 

logging interval will be 10 minutes from the last burst 

logging point, or 9:22 in this case. If burst logging had not 

occurred, the next data point would have been at 9:15. 

 

A New Interval event is created each time the logger 

enters or exits burst logging mode. See 

Logger Events

 for 

details on plotting and viewing the event. In addition, if 

the logger is stopped with a button push while in burst 

logging mode, then a New Interval event is automatically 

logged and the burst condition is cleared, even if the 

actual high or low condition has not cleared. 

The logger 

will check the high and low conditions when logging 

resumes (if the logger was configured with Allow Button 

Restart selected)

Statistics Logging 

During fixed interval logging, the logger records data for 

enabled sensors and/or selected statistics at the logging 

interval selected. Statistics are calculated at a sampling rate you 

specify with the results for the sampling period recorded at 

each logging interval. The following statistics can be logged for 

each sensor: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 

values. 

For example, a logger is configured with both the temperature 

and RH sensors enabled, and the logging interval set to 5 

minutes. The logging mode is set to fixed interval logging with 

Normal and all four statistics enabled and with a statistics 

sampling interval of 30 seconds. Once logging begins, the logger 

will measure and record the actual temperature and RH sensor 

values every 5 minutes. In addition, the logger will take a 

temperature and RH sample every 30 seconds and temporarily 

store them in memory. The logger will then calculate the 

maximum, minimum, average, and standard deviation using the 

samples gathered over the previous 5-minute period and log 

the resulting values. When reading out the logger, this would 

result in 10 data series (not including any derived series, such 

as dew point): two sensor series (with temperature and RH 

data logged every 5 minutes) plus eight maximum, minimum, 

average, and standard deviation series (four for temperature 

and four for RH with values calculated and logged every 5 

minutes based on the 30-second sampling). 
To log statistics: 
1.

 

Tap the HOBOs icon and tap the logger to connect to it. If 

the logger was configured with Power Saving Mode 

enabled, press the button on the logger to wake it up. 

When working with multiple loggers, you can also press the 

button on the logger to bring it to the top of the list. 

2.

 

Once connected, tap Configure. 

3.

 

Tap Logging Mode and then select Fixed Interval Logging. 

4.

 

Select Normal to record the current reading for each 

enabled sensor at the logging interval shown at the top of 

the screen. Do not select this if you only want to log 

statistics. 

5.

 

Select the statistics you want the logger to record at each 

logging interval: Maximum, Minimum, Average, and 

Standard Deviation (average is automatically enabled when 

selecting Standard Deviation). Statistics will be logged for all 

enabled sensors. In addition, the more statistics you record, 

Summary of Contents for MX2303

Page 1: ...ernal temperature sensor xx minutes in air moving 1 m sec MX2303 and MX2304 external sensors xx minutes in air moving 1 m sec xx seconds in stirred water Stability Drift 0 01 C 0 018 F per year Relati...

Page 2: ...ll be illuminated for 4 seconds Status LED This LED blinks blue every 4 seconds when the logger is logging unless Show LED is disabled as described in Configuring the Logger If the logger is waiting t...

Page 3: ...n on the logger for 10 seconds to reset a password Update Firmware When new logger firmware is available this action appears in the list Select it and follow the instructions on the screen Note that i...

Page 4: ...then tap Done For example select 30 days if you want the logger to log data for 30 days after logging begins d Tap Done in the Stop Logging screen 9 Select the sensor measurement types that will be lo...

Page 5: ...r clear when the sensor reading is within the logger specifications of 0 04 C resolution This means the value that triggers the alarm may differ slightly than the value entered For example if the High...

Page 6: ...r is stopped with a button push while in burst logging mode then a New Interval event is automatically logged and the burst condition is cleared even if the actual high or low condition has not cleare...

Page 7: ...ect to it If the logger was configured with Power Saving Mode enabled press the button on the logger to wake it up When working with multiple loggers you can also press the button on the logger to bri...

Page 8: ...y life is 1 year typical with a logging interval of 1 minute but may be extended to 2 years when the logger is configured with Power Saving Mode enabled Expected battery life varies based on the ambie...

Page 9: ...nect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This device complies with Part 1...

Reviews: