Hobo MicroRX Station RX2101 Manual Download Page 14

 

HOBO MicroRX Station Manual 

 

1-800-LOGGERS 14 

www.onsetcomp.com 

4.

 

For Battery Low and Sensor Failure alarms: 

a.

 

Under Device, select the Battery Low and/or Sensor 
Failure checkboxes. 

b.

 

Select how you want to be notified when these alarms 
trip: by email or text. Enter the appropriate addresses 
and then select “Send on Clear Also” if you want an 
email or text when these alarms clear as well. 

5.

 

Click Add Action if you want multiple actions to be taken 
when the alarm trips (for example send an email and a 
text). 

6.

 

Click Save. Changes will take effect the next time the station 
connects to HOBOlink. 

Red alarm symbols will appear in HOBOlink when these alarms 
trip (if enabled). 

Sensor Alarms 

To add a sensor alarm: 

1.

 

In HOBOlink, click Devices and then RX Devices, and find the 
station you want to configure. Click the arrow next to 

 

and select Alarm Configuration. 

2.

 

Click Add a Sensor Alarm. 

3.

 

Set up the Sensor Condition for the alarm. 

a.

 

Select the sensor. 

b.

 

For rain gauge sensors:

 This is based on accumulated 

rainfall in inches or mm (depending on the units set in 
HOBOlink) over a period of minutes or hours (up to 24). 
Enter the number of minutes or hours you want the 
rainfall to be accumulated over, and the amount of rain 
in inches or mm you want to trigger the alarm. 

c.

 

For all other sensors and channels:

 Select whether the 

alarm should trip above or below a value or outside a 
range. Enter the sensor reading(s) for the alarm 
threshold. Enter the number of logged data points you 
want the station to record before the alarm trips. 

Note:

 

If you are setting up a sensor alarm for a water flow 
channel that is using a stage-discharge table, be sure the 
alarm limits are within the range of the stage-discharge 
values you are using. 

d.

 

If you selected the alarm to trip above or below a 
specific reading, then select when the alarm should 
clear: above or below the same value or a different 
value. Enter the value if necessary.  

4.

 

Select the action to be taken when the alarm trips: send an 
email or text. Enter the details and then select “Send on 
Clear Also” if you want an email or text when the alarm 
clears as well. 

Important:

 Standard data fees and text messaging rates 

may apply when using text notifications. Onset does not 
charge a fee or guarantee delivery of text alerts, which is 
subject to your carrier’s service and location. See the 
HOBOlink Help for additional details on alarm 
notifications. 

5.

 

Click Add Action if you want multiple actions to be taken 
when the alarm trips (for example, send an email and a 
text). 

6.

 

Add any optional notes for this alarm. 

7.

 

Click Save. Changes will take effect the next time the station 
connects to HOBOlink. 

8.

 

Repeat steps 2 through 7 for each additional sensor alarm 
you want to add. 

Note:

 There is a limit of eight alarms 

allowed for all water level sensor channels (this includes 
barometric pressure, water pressure, differential pressure, 
water temperature, water level, and water flow). 

If an alarm trips on a smart sensor or water level sensor 
channel, the station will automatically connect to HOBOlink to 
report the tripped alarm. A red alarm symbol appears next to 
that sensor in HOBOlink when it trips. An alarm symbol will also 
appear on the LCD. 

 

Note that for water level and flow alarms, HOBOlink calculates 
the corresponding differential pressure value so that this can 
trigger an alarm by the station the same as other sensors, and 
trigger an immediate connection if there is an alarm condition. 

Alarms for HOBOlink calculated channels (such as dew point) 
are checked at the connection interval for the station. This is 
because calculated channels are generated from data uploaded 
at each connection interval. 

Accumulated rain alarms are triggered by the station as soon as 
the alarm conditions are met. The alarm will remain tripped for 
one accumulation interval after the trip. For example, if you set 
the alarm to trip for 2 inches of rainfall accumulated in 4 hours, 
then it will be 4 hours before an alarm clear message is sent 
after the alarm is triggered--unless enough rain has continued 
to accumulate that causes the alarm to trip again. 

Setting up Water Level and Water Flow 
Channels in HOBOlink (RX2103 and  
RX2104 Models) 

RX2103 and RX2104 models support a water level sensor. 
When installed, the station will automatically log four 
measurement channels: 

 

Barometric pressure 

 

Water pressure 

 

Differential pressure 

 

Water temperature 

In addition, you can set up two additional channels in 
HOBOlink: 

 

Water level 

 

Water flow 

These derived channels are only available once enabled in 
HOBOlink. The data for these two channels are calculated at 
each logging interval based on the measurements from the 
pressure and temperature channels and the settings and values 
you enter in HOBOlink. 

Tripped smart 
sensor and 
water level 
sensor alarms 

Summary of Contents for MicroRX Station RX2101

Page 1: ...X2101 and RX2103 6 AA 1 5 V lithium batteries or AC power adapter P AC 1 Battery Life RX2102 and RX2104 Typical 3 5 years when operated in the temperature range 20 to 40 C 4 to 104 F operation outside this range will reduce the battery service life Maximum connection rates with built in solar panel in full sun 10 minute connections year round for latitudes less than 40 10 minute connections throug...

Page 2: ... QIPPLS62 W IC ID 7830A PLS62W Water Level Sensor Pressure Absolute and Water Level Measurements MX2001 01 S and MX2001 01 Ti S Operation Range 0 to 207 kPa 0 to 30 psia approximately 0 to 9 m 0 to 30 ft of water depth at sea level or 0 to 12 m 0 to 40 ft of water at 3 000 m 10 000 ft of altitude Factory Calibrated Range 69 to 207 kPa 10 to 30 psia 0 to 40 C 32 to 104 F Burst Pressure 310 kPa 45 p...

Page 3: ... 005 ft water Pressure Response Time 90 1 second at a stable temperature Water Level Sensor and Cable Dimensions Sensor MX2001 0x S and MX2001 0x Ti S 2 54 cm 1 0 inches diameter 9 91 cm 3 9 inches length Cable CABLE RWL xxx 0 47 cm 0 03 0 185 inches 0 01 diameter 0 2 to 400 m 0 65 to 1 312 ft length 5 44 x 2 46 x 1 45 cm 2 14 x 0 97 x 0 57 inches connector station end 5 0 x 2 5 cm 1 97 x 0 98 inc...

Page 4: ...8 3 Mount the station 8 4 Remove the cable channel Make a note of how the cable channel is oriented when removing it 8 5 Install the water level sensor if applicable RX2103 and RX2104 models 8 6 Plug in smart sensors if applicable 9 7 Grease and reinstall the cable channel 9 8 Plug in the battery and wait for the station to connect to HOBOlink 10 9 Configure the station in HOBOlink 10 10 Start log...

Page 5: ...se this port to plug in an AC adapter see both Battery Information sections Cable Channel Use this for routing sensor cables and other wires to create a weatherproof seal Any open holes should be sealed with the integrated plugs see Setting up the Station and Deployment Guidelines Smart Sensor Connectors Use these input jacks to connect up to 5 smart sensors see Setting up the Station The station ...

Page 6: ...and other items This indicates the station is connected to a USB cable This shows the strength of the cellular signal the more bars there are the stronger the signal This will blink while connecting to HOBOlink or When the station is attempting to connect or is currently connected to HOBOlink Connection flashes on the LCD After the connection is complete Last Connection remains illuminated and one...

Page 7: ...4 representing barometric pressure water pressure differential pressure and water temperature Note that water level and water flow channels are derived in HOBOlink and not included in the channel count shown on the station LCD This will blink in the lower right part of the LCD when a firmware update is underway It will display which module or element is being updated This is a numerical code that ...

Page 8: ...p U bolts to attach the logger to a tripod or mast this is the recommended method for mounting on a mast Do not use U bolts without the saddle clamps as that could bend the mounting tabs and damage the housing or compromise the weatherproof seal The flat portion of the saddle clamps should be against the mounting tabs Use the included cable ties with the two sets of inner holes to affix the logger...

Page 9: ...r connector 7 Grease and reinstall the cable channel a Use the integrated plugs to fill any unused holes Bend the plugs up so that you can push them into the holes Once a plug is partially pushed through you can pull on the part of the plug that is inside the case You may need to bend the ends of the channel slightly to widen the holes for installing the plugs b Lightly coat the portion of the sen...

Page 10: ...een to the next or use the left menu to select a specific item to configure Follow the steps in the next subsections to configure the readout settings smart sensors if applicable and water level sensor if applicable Note Click Save or Next in any screen to save your changes You will lose any changes made if you click Back without clicking Next or Save first Important Do not configure the water lev...

Page 11: ...checkbox and fill in the Scaled Units Multiplier Offset and Scaled Measurement Type fields h Click Save You can also click Next to move from one smart sensor to the next and save the sensor configuration Clicking Back does not save the configuration changes i Repeat steps e h for any additional smart sensors you need to configure Water Level Sensor Module Configuration You can configure both the g...

Page 12: ...ake a reference level reading measuring the water level from your reference point Important Note the reference level reading as well as the date and time it was taken 12 Configure the water level and water flow channels in HOBOlink RX2103 and RX2104 models Perform the following steps in the field in HOBOlink with a mobile device to verify that the system is logging the water level correctly while ...

Page 13: ...pe the name of the delivery schedule and the frequency of delivery Enable the Active checkbox Select other settings if desired 4 Under Select Data to Export choose the name of the custom data export you want to be delivered or follow the previous set of steps to set up a new data export 5 Under Data Destination select FTP SFTP or Email for the delivery method and fill in the appropriate fields 6 C...

Page 14: ... your carrier s service and location See the HOBOlink Help for additional details on alarm notifications 5 Click Add Action if you want multiple actions to be taken when the alarm trips for example send an email and a text 6 Add any optional notes for this alarm 7 Click Save Changes will take effect the next time the station connects to HOBOlink 8 Repeat steps 2 through 7 for each additional senso...

Page 15: ...calculated starting with the next connection to HOBOlink Note that the reference water level information entered in this step will not affect any previously logged data It will only be use for data logged from the point of the next connection to HOBOlink forward If this is the first time that water level has been configured for this station then the data stored will go back to the date and time of...

Page 16: ...r If the reference point is below the water level then enter a positive number 8 Click Save Water flow will be calculated starting with the next connection to HOBOlink Tip It is good practice to set an alarm for the water level just below the top of the v notch so you know when the flow data is close to exceeding its valid range See Setting System and Sensor Alarms Setting up a Water Flow Channel ...

Page 17: ... in the second diagram The distance from the water level reference point to the notch crest which is represented as WLc in the second diagram The HOBOlink water flow calculations for a trapezoidal weir assume the following The slopes of the notch sides must be 4 vertical change horizontal change Head height h should be measured at a distance of at least 4 x hmax upstream of the weir The weir shoul...

Page 18: ...he Flume Coefficient may also be labeled K by some manufacturers so be sure to check how the coefficient is being used in their equation to make sure it is the right one to enter here 7 For the Head Offset enter the value in either meters or feet using the same units as you entered the reference water level in the water level channel An example of Head Offset is shown in the following diagram of a...

Page 19: ...hose as well For example when a stream overflows its banks the stage discharge relationship will change dramatically The curve fit equation will no longer be valid so you will need a different curve fit equation for stage levels in this range Important Make sure the station has started logging and you have configured the water level channel with a reference level reading before performing these st...

Page 20: ... up the channel count on the LCD should be updated If you removed the sensor the total channel count will decrease by three If you added the sensor the channel count will increase by three Barometric pressure is always included in the channel count whether the sensor is plugged in or not If you added a water level sensor be sure to obtain a reference water level reading and configure the water lev...

Page 21: ...he other holes in the cable channel and can accommodate both the AC adapter cable and the integrated plug at the same time Do not store excess sensor cable wire coiled inside the station case or within one foot outside the case Protect cables and wires with conduit Exposed cables can be chewed by rodents Make sure the total cable length for all installed smart sensors does not exceed 100 m 328 ft ...

Page 22: ...ge to the cable Installing the Grounding Wire If you are using a grounding wire CABLE MICRO G attach it to the grounding wire port on the back of the station Use the screw and washer included with the grounding wire to attach it to the port If you are mounting the station on a tripod or mast use the optional U bolts U BOLT KIT2 Attach the grounding wire under one of the nuts on one end of the U bo...

Page 23: ...atteries may rupture or explode Do not dispose of the logger or batteries in fire Do not expose the contents of the batteries to water Dispose of the batteries according to local regulations for lithium batteries To replace the batteries 1 Open the station door 2 Disconnect the battery cable 3 Use a flat head screwdriver to remove the screw below the battery cover 4 Pull to remove the battery cove...

Page 24: ...d screwdriver to secure the cover with the screw 8 Plug in the battery cable 9 Make sure the built in solar panel cable is plugged in If you are using an external solar panel make sure the built in solar panel cable is tucked inside the station door Plug in the external solar panel Lightly coat the portion of the cable that will be placed in the cable channel with a small amount of silicone grease...

Page 25: ...er Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interfe...

Reviews: