VERTIC 2 Fortissimo user guide 

Version 12/2018  

Summary of Contents for VERTIC 2 Fortissimo

Page 1: ...1 VERTIC 2 Fortissimo user guide Version 12 2018 2 2 F Fo or rt ti is ss si im mo o 330Kg 52 stones ...

Page 2: ...2 VERTIC 2 Fortissimo user guide Version 12 2018 ...

Page 3: ...de anytime TABLE OF CONTENTS 1 PREAMBLE page 4 2 DESCRIPTION page 4 7 2 1 Where to use Vertic 2 Fortissimo 2 2 General description 2 3 Technical description 2 4 Dimensional characteristics 2 5 Labelling 3 VERTIC 2 FORTISSIMO ASSEMBLING page 8 3 1 Package contents first use 3 2 Vertic 2 Fortissimo assembling RECOMMENDED CHECKS page 9 EVALUATION OF THE NUMBER OF HOIST OPERATING CYCLES page 10 MAINTE...

Page 4: ...use of this product This information has been carefully checked by HMS VILGO This is not an exhaustive list and HMS VILGO is not responsible for any kind of mistakes or omissions 2 DESCRIPTION 2 1 WHERE TO USE VERTIC 2 FORTISSIMO The Vertic 2 Fortissimo is a lifting and transfer device It is used to transfer disabled or elderly people At home to transfer the patient to an armchair a toilet seat or...

Page 5: ...n classification Class II Applied parts Type B Electrical equipment conformed to NF EN 60 601 1 and to NF EN 60 601 1 2 Duty cycle 2 min 18 min 10 Input 230V 50 HZ Actuator input 24 V CC Battery charge required time 15 hrs Battery autonomy 40 cycles Electrical equipment protection remote control IP X6 control box IPX4 actuator IP X4 Side foot Electric actuator Mast Control box Driving handles Arms...

Page 6: ...mum back height 6 5 in Total lift lifting radius overall dimension 1 Lifting radius at the highest suspension point 3 5 in 2 Total lift 3 4 30 in 3 Minimal suspension point height 31 in 4 Lifting radius at lowest point 17 7 in 5 distance between floor hooks in the low position 29 9 in 6 distance between floor hooks in the high position 57 5 in 7 Overall height when in the low position 42 3 in 8 Ov...

Page 7: ... up of the medical disposal It also gives to you the security loading charge for a good use There are 4 labels One on the basis of the Vertic 2 Fortissimo one on the electric actuator one on the control box and one on the remote control Standards By this label HMS VILGO ensures that Vertic 2 Fortissimo hoist is conformed to European directive 89 336 referring to electromagnetic compatibility and t...

Page 8: ...fasteners and drop on the ground the Vertic 2 Fortissimo All connections are made in factory and the battery delivered with it is fully charged so that you can start using the hoist immediately Check the emergency stop of control box is unlock and try it without a patient to get familiar with its use and functions Main cable x1 1 pocket contening User guide x1 Vertic 2 Fortissimo ...

Page 9: ... Actuator arms connection Check fixing screw wear level 10 Hooks check welding of the 3 hooks 11 Emergency release Check electrical and manual release operating 12 Sling Check seams 13 Leg rest check the welding point of the leg rest check the blocking device 14 Pole pole support connection Check knobs and screws presence and assembling Check supports poles welding 15 Foot rest wheels Check the co...

Page 10: ... 40880 45990 51100 15 5475 10950 16425 21900 27375 32850 38325 43800 49275 54750 16 5840 11680 17520 23360 29200 35040 40880 46720 52560 58400 17 6205 12410 18615 24820 31025 37230 43435 49640 55845 62050 18 6570 13140 19710 26280 32850 39420 45990 52560 59130 65700 19 6935 13870 20805 27740 34675 41610 48545 55480 62415 69350 20 7300 14600 21900 29200 36500 43800 51100 58400 65700 73000 21 7665 1...

Page 11: ...2 Fortissimo user guide Version 12 2018 M MA AI IN NT TE EN NA AN NC CE E B BO OO OK K HOIST REFERENCE Serial number Date of 1st use DATE Number of cycles evaluated MAINTENANCE AND QUALITY CHECKING NAME APPROVALS ...

Page 12: ...on HH 12x60 screw Ecrou frein Ø12 A501249 A001080 1 actuator mast connection Ø 40 14 2 e2 brass ring Ø14 M12 Lg91 axle Stop nut Ø 12 Ø12 ZI washer A501404 A501402 A001080 A001088 4 2 4 6 mast arms connection HH 10x100 screw Ø10 stop nut D16 di 10 lg86 bushing D22 di 16 lg60 bushing A501375 A001072 A501406 A501409 2 2 2 2 Side foot cross piece connection 3in twin wheel under foot rest HH 10x30 ZI s...

Page 13: ... foot Electric actuator entry n 2 Main cable connection Actuator connection réf A501530 actuator cable control box CBJ connection réf A501451 Locking clip réf A501491 Check locking clip set up 4 USE 4 1 PLUGGING Plugging to CBJC control box Locking clip set up Remote control connection ...

Page 14: ...r close side foot to get closer to the patient 2 Adjust the leg rest up or down to block the patient s shins against it You can also adjust the incline 3 Lock brakes 4 Set up the sling 5 Press the up button on the handset to lift up the main pole make sure the sling is not stuck and does not injure the patient while doing this 6 Unlock brakes to transfer the patient 7 Press the down button to brin...

Page 15: ...function can be used if the electrical system is not efficient 4 4 CONTROL BOX OPERATING BATTERY CHARGE If the charging level is too weak an alarm sounds in a constant way when the control box is activated The charging of the battery is then necessary The control box has its own integrated charger To charge batteries connect the control box to the 220 VAC main power line until the green led turn o...

Page 16: ...ng by looking at its seams condition immediately replace a sling that starts to show any tear Electrical parts Test the emergency stop switch while lifting and lowering the support beam Look at the electrical cables cover Test crushing safety system by placing an item on the support beam going down way Chassis Check the brake of the back wheels Look at the global geometry on the Vertic 2 Fortissim...

Page 17: ...NY RETURN WITHOUT SERIAL NUMBER LABEL WILL NOT BE COVERED BY THE WARRANTY The warranty does not cover any breakdown due to inappropriate or wrong use or maintenance in spite of the specifications and use instructions stated in this manual electrical shock torn cables etc Warranty does not cover any breakdown or malfunction due to parts being swapped over or replaced with used or unapproved parts 5...

Page 18: ...18 VERTIC 2 Fortissimo user guide Version 12 2018 ...

Page 19: ...19 VERTIC 2 Fortissimo user guide Version 12 2018 ...

Page 20: ...s vilgo com Site de Lille Parc Eurasanté 243 Rue Allendé 59120 LOOS LEZ LILLE Tél 33 0 3 20 84 81 50 Fax 33 0 3 20 59 58 08 Site de Bergerac Rte de Sainte Alvère Creysse BP 212 24102 BERGERAC Cedex Tel 33 0 5 53 74 45 50 Fax 33 0 5 53 63 06 07 2 2 F Fo or rt ti is ss si im mo o ...