USER’S GUIDE

PANEL CUT-OUT DIAGRAM

M

H

D I G I T A L

®

PS-100

PANEL MOUNT TDS CONTROLLER

1. Using a knife, cut the diagram out (cut on the outer part of the line).

2. Align the cut-out to your panel and draw cut marks.

3. Cut the hole in the panel to the precise dimensions of the cut-out:

     2-11/16 in. x 2-11/16 in. (68 mm x 68 mm)

 

--> 

See the installation section for complete instructions.

CONTACT DIAGRAM

11

/

16 

in. (68 mm)

 

11

/

16 

in. (68 mm)

 

8

9

10

11

12

13

14

1

2

3

4

5

6

7

  AC

N.C.

N.O.

WHITE

RED

BLACK

ALARM

(BUZZ)

SENSOR

110-220V

RELAY (SWITCH)

Summary of Contents for PS-100

Page 1: ...lign the cut out to your panel and draw cut marks 3 Cut the hole in the panel to the precise dimensions of the cut out 2 11 16 in x 2 11 16 in 68 mm x 68 mm See the installation section for complete instructions CONTACT DIAGRAM 2 11 16 in 68 mm 2 11 16 in 68 mm 8 9 10 11 12 13 14 1 2 3 4 5 6 7 AC N C N O WHITE RED BLACK ALARM BUZZ SENSOR 110 220V RELAY SWITCH ...

Page 2: ...3 Sensor cable grey 5 Mounting brackets 2 Sensor SP 1 4 Power cord black 6 U S plug adapter TDS Range 0 999 ppm mg L Resolution 1 ppm Accuracy 2 of the reading Temperature Compensation Automatic ATC 1 60oC Calibration Manual by trimmer pot calibration screw Set Point Single point controlled by on screen up down buttons to any point within the range Set Point Relay single isolated 2A Max 220V resis...

Page 3: ... Any consequential damages incidental damages or incidental expenses including damages to property This includes damages from abuse or improper maintenance such as tampering wear and tear water damage or any other physical damage The Company s products are not waterproof and should not be fully submerged in water Products with any evidence of such damage will not be repaired or replaced See additi...

Page 4: ...eeds The calibration solution should be NaCl HM Digital s 342 ppm NaCl is recommended 2 Check the controller against the calibration solution Prior to inserting the sensor into the calibration solution be sure the sensor is dry and free of any TDS residue If the reading on the controller does not match the calibration value you will need to re calibrate If using a fitting calibrate with the fittin...