Force gauge         Hand Terminal HT 6.0

7

Operation

The menu is used via the four buttons:
- It is opened with the 

MENU

 button.

- You can scroll through the menu options / submenu options with the 

 buttons, or increase or decrease adjustable values.

- The 

enter

 button is for opening the selected menu option or to confirm the entry.

- The menu is closed with the 

escape

 button. The device is then in measuring mode.

4.2.1.  “About” menu option

With this menu option, you can read out the following information about your device:

Firmware

Version of the installed software

Data last cal.

Date of the last calibration

Serial No.

Serial number

Procedure

:

 - Press 

MENU

 to open the menu

 - Open menu item “1. About” with 

enter

 - Select the desired submenu option with 

 and open it with enter

 - The device-specific information can be read in the display
 - Return to the submenu with 

enter

 - Select another submenu option with 

or

 - Select the submenu option “1.4 Back” with 

 and return to the main menu with 

enter

or

 - Exit the menu with 

escape

. The device returns to measuring mode.

4.2.2.  “Configuration” menu option

With this menu option, you can configure the following settings:

Auto off

Select the time after which the device automatically switches off to save batteries.
You can select one of the following: off, 60, 30, 8 min. (Factory setting: 60 min)

Averaging

Here, you can select the desired measuring rate customised for your application, at the same time, the 
frequency also changes with which the measured value display is updated in the display. The lower 
measuring rate is suited to slower loads, the higher rate is suited to short, fast load changes. The higher 
the measuring rate, the better the detection of peak loads. However, with the high measuring rate the 
measured value display becomes slightly more disturbed, as the noise element increases.
You can select one of the following: 
Setting

Measuring rate

Display frequency

slow

10 Hz

2 Hz

middle

20 Hz

5 Hz

(Factory setting)

fast

100 Hz

10 Hz

Contrast

Here you can set the contrast in the display, as you can best read it.
You can select between 10 - 100 % in 10%-steps

HKM-Messtechnik GmbH

www.hkm-messtechnik.com

Operating Instructions Force gauge

Status 03.2011

Summary of Contents for HT6.0

Page 1: ...Force gauge Sensor Hand Terminal HT6 0 Operating Instructions All rights reserved by HKM Messtechnik GmbH 79110 Freiburg Germany ...

Page 2: ...tion 6 4 1 Switching on 6 4 2 Using the menu MENU 6 4 2 1 About menu option 7 4 2 2 Configuration menu option 7 4 3 Measuring 8 4 3 1 Zero positioning and taring 9 4 3 2 Switching the measured value display PEAK 9 4 3 3 Performing the measurement 9 4 4 Changing the battery 10 4 5 Turning off 10 5 Calibration cycle 10 6 Declaration of conformity 11 HKM Messtechnik GmbH www hkm messtechnik com Opera...

Page 3: ...y only be loaded up to its nominal load The nominal load can be found on the type plate of the transducer You can also find this information in the declaration of conformity on page 11 or read it on the force gauge s display when switching it on There must be no transverse forces affecting the transducer The hand held terminal must not be subject to any forces This force gauge is only intended for...

Page 4: ...6 Bit Hand held terminal s power supply 4 battery type AA Operating time 8h continuous operation Transducer s power supply 3 3V Transducer s bridge resistance 100Ω to 1kΩ Auto off time 8 30 60 min off Operating temperature range 0 C to 50 C Storage temperature range 20 C to 70 C Protection class of the hand terminal IP40 Material and dimensions of the hand terminal ABS 78mm 145mm 44mm b l h HKM Me...

Page 5: ...Name in the menu Function in the menu light blue font ZERO POSITIONING TARING Zero positioning taring and Clear peak values up scroll up in the menu increase numerical values PEAK Toggle between displays of the current measured value of the maximum value largest value in positive direction of force of the minimum value largest absolute value in negative direction of force down scroll down in the m...

Page 6: ...a previously tared value has been retained The peak values were deleted 4 2 Using the menu MENU The menu provides the following options Read out information about the device About menu option Configure device settings Configuration menu option Calibrate the device Service menu option this area is password protected and can only be configured by the factory Enter Enter Enter Enter Enter Esc Esc Esc...

Page 7: ...menu with escape The device returns to measuring mode 4 2 2 Configuration menu option With this menu option you can configure the following settings Auto off Select the time after which the device automatically switches off to save batteries You can select one of the following off 60 30 8 min Factory setting 60min Averaging Here you can select the desired measuring rate customised for your applica...

Page 8: ...eady Fast or unsteady load suspension results in dynamic load peaks which can lead to overload damages up to breaking of the load suspension devices or measuring device In general the applied load consisting of static dynamic tared load must not exceed the nominal load of the transducer Getting under a suspended load or entering the direct risk area is prohibited Information about the nominal load...

Page 9: ...the PEAK button In this setting the measured value display corresponds to a drag indicator which is further advanced in case of increasing decreasing values If you press the PEAK button once the maximum value is displayed in positive load direction Peak Max is displayed below the measured value Press the PEAK button once more to display the minimum value i e the largest absolute value in negative ...

Page 10: ...he nose click into place in the housing by applying slight pressure Thread the housing first on the cable side and then on the opposite side back into the protective cover 4 5 Turning off By holding the button down the force gauge is switched off The peak values are deleted If the system was tared this tare value is kept even if the force gauge is switched on or off The individual settings you ent...

Page 11: ... Electro magnetic fields Electro magnetic fields mobile radiation In case the above mentioned product has been changed without having obtained permission from HKM Messtechnik GmbH this declaration is no longer valid Freiburg 15 February 2011 Dipl Ing FH Martin Schlegel Plant management HKM Messtechnik GmbH The content of this print including excerpts must not be passed on to third parties without ...

Reviews: