![hivolt HA05B2 Operating Manual Download Page 16](http://html1.mh-extra.com/html/hivolt/ha05b2/ha05b2_operating-manual_2140416016.webp)
16
hivolt.de
GmbH & Co. KG
Oehleckerring
40
∙
D-22419
Hamburg
∙
Germany
∙
+49 40 537122-0
∙
+49 40 537122-99
∙
∙
www.hivolt.de
5.11
External Connectors
Signals: BNC
HV Outputs:
SHV
Power line:
IEC320-1 (HAR12S)
Earth:
Bolt M4 (HAR12S)
Power supply:
2 pole Phoenix Combicon Connector (HAR2, HAR2-W)
Mating connector:
2 pole Phoenix Combicon Plug MSTB 2.5/2-ST-5.08
5.12
Internal Connectors
Internal connectors for HA05B2 plug-in modules (DIN41612, Type F)
Pins
Signal
D2 B2 Z2
+VHV (+540V)
D4 B4 Z4 GND
D6 B6 Z6
VO (Amplifier output)
D8 B8 Z8 GND
D10 B10 Z10 -VHV (-540V)
D12 B12 Z12 GND
D14 B14 Z14 n.c.
D16 B16 Z16 n.c.
D18 B18 Z18 Internal connection, do not connect
D20 B20 Z20 Internal connection, do not connect
D22 B22 Z22 Internal connection, do not connect
D24 B24 Z24 Internal connection, do not connect
D26 B26 Z26 Internal connection, do not connect
D28 B28 Z28 Internal connection, do not connect
D30 B30 Z30 +VS (+24V)
D32 B32 Z32 GND
5.13
Grounding
The M4 ground bolt on the back panel of the device must be connected to the central grounding point
of the test setup and to protective earth (
HAR12S
).