5-12-2. STERRAD
®
sterilization
Sterile conditions of applicable sterilization methods are as follows.
The applicable gas is listed below.
General name
Trade name
Manufacturer
Hydrogen peroxide
(58% density)
STERRAD
®
Sterilization system
(STERRAD
®
50, 100S, 200,
NX or 100NX )
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Regarding the operation of the sterilizer, refer to the documentation supplied with the sterilizer.
Remark:
Some discoloration of the probe may occur, but this does not affect performance or safety.
Warning
Do not use the waterproof cover during sterilization process
.
Caution
Do not sterilize the probe using the STERRAD system if the probe is not compatible with the
STERRAD system. STERRAD compatibility is shown by the STERRAD label on the con-
nector.
Perform STERRAD sterilization only for STERRAD compatible probes, otherwise it can
cause damage or deterioration to the probe.
Do not put the probe directly into the sterilization pouch*.
Otherwise the pouch sticks to the cable and results in damage to the cable. Completely wrap
the entire probe (including the probe tip, cable and connector) with sterilization wraps* before
putting it into the sterilization pouch*.
*: A Johnson & Johnson company Division of Ethicon, Inc. product
* STERRAD
label
Summary of Contents for UST-5550
Page 2: ...MN1 5575 Rev 11 ii ...
Page 14: ... 8 MN1 5575 Rev 11 2 Storage case Label A Label C Label B Label D ...
Page 16: ... 10 MN1 5575 Rev 11 ...
Page 22: ... 16 MN1 5575 Rev 11 ...
Page 50: ... 44 MN1 5575 Rev 11 ...
Page 52: ... 46 MN1 5575 Rev 11 ...
Page 54: ... 48 MN1 5575 Rev 11 ...
Page 58: ... 52 MN1 5575 Rev 11 ...