-4-

MN1-5845 Rev. 4

2-2. Name of each parts

The name of each part is shown in Figure 2 and the explanation for each part is listed in Table 1.

Orientation mark

Connector

Ultrasonic radiation part

Cable

Probe tip

Figure 2 Name of each parts

Table 1  Name of each part and its explanation

Name

Explanation

Ultrasonic radiation part

Ultrasound is radiated from this part. The electronic 

sector transducer is integrated underneath this part.

Orientation mark

The side of the orientation mark corresponds to the side 

of the orientation mark on the image.

Probe tip

This part is gripped during operation.

Cable

Cable transfers electric input/output signals.

Connector

The connector is the part which is connected to the 

ultrasound diagnostic instrument.

 Caution

Do not pull, bend, twist, or apply excessive force to the cable.

The probe may malfunction due to cable disconnection. 

Do not subject the ultrasonic radiation part to hard impact.

The impact may cause damage to the transducer, and that results in noise or no echo in the image.  

In most cases, the ultrasonic radiation part itself is not damaged because the part is made of elastic 

material. 

Summary of Contents for S12

Page 1: ...ad through this Instruction Manual as well as the separate Instruction Manual Safety MN1 5980 and Cleaning Disinfection and Sterilization MN1 5998 carefully prior to use Keep this Instruction Manual together with the ultrasound diagnostic instru ment for any future reference Hitachi Ltd 2013 2017 All rights reserved MN1 5845 Rev 4 Q1E EP1471 ...

Page 2: ...ii MN1 5845 Rev 4 ...

Page 3: ...ds and injuries to the operator or patient Danger Indicates an imminently hazardous situation which if not avoided will result in death or serious injury to the operator or patient Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury to the operator or patient Caution Indicates a potentially hazardous situation which if not avoided may re...

Page 4: ...ral Information 1 1 1 Intended use 1 1 2 Classification of ME equipment 1 1 3 Standard components 1 2 Specifications and Parts name 3 2 1 Specifications 3 2 2 Name of each parts 4 3 Preparations before use 5 3 1 Visual check 5 3 2 Confirmation of cleaning disinfection and sterilization 5 3 3 Operation check 5 4 Operation 7 ...

Page 5: ...ies to the operator or the patient 1 2 Classification of ME equipment This probe is classified as follows according to IEC60601 1 Please refer to the section 2 1 for the applied part the part treated as the applied part and the range of IPX7 Classification based on the degree of protection against electric shock Type BF applied part Classification for protection against ingress of liquids IPX7 Wat...

Page 6: ... 2 MN1 5845 Rev 4 ...

Page 7: ...r each country please refer to the instrument instruction manual or contact your local Hitachi representative Field of view 90 Frequency 3 0 MHz Cable length 2 2 m Service life 3 years Applied part Probe tip including ultrasonic radiation part see the section 2 2 Part treated as Applied part Cable up to 1 m length from the probe tip IPX7 range See Figure 1 Measurement accuracy Refer to the instruc...

Page 8: ...e orientation mark corresponds to the side of the orientation mark on the image Probe tip This part is gripped during operation Cable Cable transfers electric input output signals Connector The connector is the part which is connected to the ultrasound diagnostic instrument Caution Do not pull bend twist or apply excessive force to the cable The probe may malfunction due to cable disconnection Do ...

Page 9: ...ect the probe to the ultrasound diagnostic instrument and check that the displayed scan type and frequency correspond to those of the probe Check also that there is no abnormality in the image Remark Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to connect the probe and information displayed on the monitor If the probe is operated in still air brightn...

Page 10: ... 6 MN1 5845 Rev 4 ...

Page 11: ...rce Use with excessive force could result in injury to the patient Scan for minimum time necessary at the lowest possible acoustic output Acoustic output may affect the patient s internal tissues For details about the acoustic output please refer to the documentation supplied with the ultrasound diagnostic instrument Do not touch the connector terminal pin of the probe Electrostatic discharge may ...

Page 12: ...08 31 Manufacturer Hitachi Ltd 2 16 1 Higashi Ueno Taito ku Tokyo 110 0015 Japan Contact 81 3 6284 3668 http www hitachi com businesses healthcare index html Overseas Offices Hitachi Medical Systems GmbH Otto von Guericke Ring 3 D 65205 Wiesbaden Germany EU Importer Hitachi Medical Systems Europe Holding AG Address Sumpfstrasse 13 CH 6300 Zug Switzerland ...

Reviews: