- 15 -
Q1E-EP1441
7.4.2
Rinsing and drying
Use sterile technique when removing the probe from high
level disinfectant solution and rinse the probe
thoroughly by soaking in sterile water for more than 10
minutes (See Fig.4). After rinsing the probe, air-dry
it completely.
(1) Remove all debris and residue of sterilant/disinfectant solution
from the surface of the probe sufficiently. Debris and residue
of sterilant/disinfectant may affect patient health.
(2) The probe connector is not waterproof. Do not allow liquid to
contact the connector.
7.5
EtO Sterilization Process
L34 is compatible with Ethylene Oxide (EtO) Sterilization. This
manual does not describe the instruction of EtO sterilizer,
therefore, follow the instructions of your sterilizer regarding
the operation.
7.5.1
EtO
Sterilization
Ensure that the probe is cleaned in accordance with the
procedure stated in “7.3 Cleaning Process” before the
sterilization. Perform sterilization operations with
the condition given in Table 7.
Table.7 EtO Sterilization condition
Preconditioning: none
Conditioning in Chamber (Dwell):
Temperature:
122.0-131.0°F (equivalent of 50-55°C)
Humidity:
40-90 % RH
Prevacuum:
1.93-3.89 PSIA (22-26" Hg, 13.26-26.73kPa)
Time: 30-45
minutes
Exposure:
Temperature:
122.0-131.0°F (equivalent of 50-55°C)
Sterilant gas:
10% E0/90% HCFC
Gas Concentration 600±30 mg/L
Exposure time:
125 minutes
Post-vacuum: 1.93-3.89
PSIA (22-26" Hg, 13.26-26. 73kPa), 2 times
Aeration:
Temperature: 53°C
Time:
11 hours and 30 minutes to 12 hours
WARNING
DM 120918 v2