Rev0 10.03.24
-G3-
5. Installation
5.1 Preparation
Conduct items shown below.
No.
Item
Procedure
1
Components Check their names and quantity with the attached component list.
2
Cleaning
If the device is contaminated, clean up it with gauze and ethanol.
3
Appearance Ensure there is no scratch, deformation, or discoloration on the device.
4
Power Supply Required Power (Voltage): 400W(AC100V)
・
200W(AC200V)
Frequency: for both 50/60Hz
Earth
:
The third-class level grounding is preferable.
(Need to confirm with customer)
5
Input Voltage
Measurement
C H
E
Voltage should be measured at the power outlet.
Voltage between (H)–(C) should be AC90 – AC110V/ AC190 – AC240.
アース工事の確認は特に注意すること。
E: Earth, H: Hot line, C: Cold line (depending on countries)
6
Hopper
Attachment
(1) Peel off the tapes from the main unit and the hopper part
(2) Pushing the Main Latch Button, open the Main Cover, and remove the Mirror
Protection Sheet and Support Plate.
(cont.)
Pay particular attention to the Grounding
tape
Support
Plate
tape
Support
Plate
Mirror Protection Sheet
Hopper Attaching Screw
Elevator Arm
Hopper Attaching screw
Top Cover
Top Latch Button
Main Latch Button
Main Cover
Side Cover