9
CP-SX1350(P5SX+)
Adjustment procedure 1
(when a color differential meter is used)
1. First adjust [MID-L] tone [G:].
2. Select adjustment point [No.2][G:].
When the background is not [G] monochrome,
press the [ENTER] key on the Remote control
transmitter to change to solid [G] monochrome.
3. Measure the illumination at adjustment points
No. 2, No.3, No.10 and No.11.
The values should be:
No.2 = Y2 [lx] No.10 = Y10 [lx]
No.3 = Y3 [lx] No.11 = Y11 [lx]
4. No.2 and No.3 adjustment point have the
average of Y2 and Y3.
Y2 = ( Y2 + Y3 ) / 2 ± 3 [%]
Y3 = ( Y2 + Y3 ) / 2 ± 3 [%]
5. No.10 and No.11 adjustment point have the
average of Y10 and Y11.
Y10 = ( Y10 + Y11 ) / 2 ± 3 [%]
Y11 = ( Y10 + Y11 ) / 2 ± 3 [%]
6. Then adjust [MID-L] tone [R] and [B].
When the background is [G] monochrome,
press the [ENTER] key on the Remote control
transmitter to change to solid white.
7. Measure the color coordinates of adjustment
point [No.1] and make a note of them.
Assume that they are x = x1, y = y1.
Note:
W h e n t h e C L - 1 0 0 c o l o r a n d c o l o r
difference meter is used, the [ ](delta)
mode is convenient. When adjustment
point [No.1] color coordinate has been
selected, set the slide switch on the side
to [ ](delta) while holding down the [F]
b u t t o n o n t h e f r o n t p a n e l . T h e
measurement shown after this displays
the deviation from measurement point 1.
8. Measure the color coordinates of measurement
point [No.2] and adjust [No.2][R:] and [B:] so
that the coordinates are as follows.
x = x1 ± 0.005 , y = y1 ± 0.010
9. Similarly, measure adjustment points [No.3] to
[No.17] and adjust their color coordinates
starting in order from the small number points.
This completes adjustments required for [MIN].
Note:
Since excessive correction may lead to a
correction data overview during internal
calculations, use the following values for
reference.
[No.2] to [No.5] ± 40 or less
[No.6] to [No.9] ± 50 or less
[No.10] to [No.13] ± 70 or less
[No.14] to [No.17] ± 120 or less
10. T h e n a d j u s t [ M I N ] t o n e [ G ] s o t h a t t h e
adjustment data set two times as much as
[MID-L] tone [G].
This completes [G] color adjustments.
11. Then adjust [MIN] tone [R] and [B].
Select [No.2] [B:] and press the [ENTER] key
on the Remote control transmitter to change to
solid white.
12. Measure the color coordinates of adjustment
point [No.1] and make a note of them.
Assume that they are x = x1, y = y1.
13. N o w m e a s u r e t h e c o l o r c o o r d i n a t e s o f
measurement point [No.2] and adjust [No.2][R:]
and [B:] so that the coordinates are as follows.
x = x1 ± 0.005 , y = y1 ± 0.010 (Target)
x = x1 ± 0.020 , y = y1 ± 0.040
14. Similarly, measure adjustment points [No.3] to
[No.17] and adjust their color coordinates
starting in order from the small number points.
This completes [MIN] tone adjustments.
15. Now make similar adjustments for [MID-H] tone.
(Adjust [MID-H] tone [G] so that the adjustment
data set half as many as [MID-L] tone [G].)
16. Now make similar adjustments for [MAX] tone.
(Adjust [MAX] tone [G] so that the adjustment
data set half as many as [MID-L] tone [G].)
Summary of Contents for CP-SX1350 P5SX+
Page 38: ...38 CP SX1350 P5SX 8 Disassembly diagram M Meter screw T Tapping screw ...
Page 39: ...39 CP SX1350 P5SX M Meter screw T Tapping screw ...
Page 53: ...53 CP SX1350 P5SX 12 Connector connection diagram Connector connection diagram ...
Page 73: ...PWB assembly SIGNAL 1 P5SX ...
Page 74: ...PWB assembly SIGNAL 2 P5SX ...
Page 75: ...PWB assembly KEYPAD P5SX ...
Page 76: ...PWB assembly DUST SENSOR P5SX ...
Page 78: ...CP SX1350 YKNo 0553E Digital Media Division QR64671 Printed in Japan JE ...