- 36 -
Q1E-EP1451
9. Specifications
9.1
Probe
Type :
C41L47RP
Probe
Acoustic working
frequency
Convex scanning
: 6.5MHz
Linear scanning
: 7.5MHz
Technology :
High
density Convex Array and Linear Array
Probe
Dimensions :
See
Fig.15
Weight :
Approx.1.3kg
(Including cable and connector)
Scanning angle (Convex) : 200 deg
Field of view (Linear) : 64mm
Probe materials
: Bio-compatible allergy free components
Acoustic output
: According to IEC 60601-2-37 (See Main Unit
manual.)
Applicable system
: Depending on production and upgrade
status
For detailed information contact a
service support.
Classification :
MDD classification
Ⅱ
a.
Cleaning
: Applicable detergents are listed in the
suppliers list
Disinfection
: Applicable disinfectants are listed in
the suppliers list
Sterilization
: ETO gas sterilization
:
Plasma
sterilization
Operating conditions:
Ambient temperature
: +10 – +40°C
Contact surface temperature
(Temperature of examinee) : max. 42°C
Relative humidity
: 30 – 75%
Storage/Transportation conditions:
Temperature
: –10 – +50°C
Relative humidity
: 10 – 90%
(Subject to no condensation)