-36-
MN1-1014 Rev. 14
5-2-2. Gas disinfection
Applicable gases
General name
Trade name
Manufacturer
Formalin gas
F. gen
(14% formaldehyde)
Aso Pharmaceutical Co., Ltd.
Disinfection procedure
Washing
Be sure to wash and fully dry the probe before disinfection.
See section 5-1.
Gas disinfection
Disinfection the following conditions:
Temperature:
Room temperature
Humidity:
Normal humidity
Atmospheric pressure: Normal pressure
Aeration
Aerate the following conditions to ensure that no gases remain.
Temperature:
10°C to 60°C
50°F to 140°F
Humidity:
30% to 85%
Compression
to +101.175kPa
Decompression
to -100.000kPa
Period:
5 days (natural aeration)
Warning
Perform full aeration after gas disinfection.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.